Dental Chair with Hopkins rod endoscope · · Motorised dental chair complete with glass reflected light Accessories include: suction machine

Quote separately for Hopkins rod endoscope (each items listed to be priced separately) o o o o o 176 mm length and 8mm diameter, 70 degree view Cold light source with SMPS power supply, 15V and 150 Watt Fiberoptic cable, 4mm / 230cm Camera for recording Interfacing software with standard desktops / laptops

Interested vendors are advised to visit the Department of Radiotherapy to seek clarifications if any in regard to the above items.

Motorised OT table with positioning accessories Remote controlled, electronically operated, electro-hydraulic operating table which is: · Three or Four sectioned, eccentrically positioned · Radio transluscent table top with appropriate sliding mechanisms to facilitate the use of a CArm image intensifier · Multiple operating positions · Standard accessories that include: leg support; knee supports; shoulder support; arm support; back support; shoulder support; head end detachable; leg end detachable; water proof mattresses; wrist slips; kidney bridge; neurosurgery head rest; cassette holders for radiographic films Floor mounted, dome shaped OT light, with a dome diameter of at-least 50 cm and at-least 5 bulbs Quote separately for knee and leg positioning devices and body vacuum cushions used to immobilize patients in supine position during radiotherapy related procedures Interested vendors are advised to visit the Department of Radiotherapy to seek clarifications if any in regard to the above items.

Annexure-III

List of instruments to be used in open surgical procedures in Uro OT S.no. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 Name of instrument Bachau's towel clamp Needle holder (Duro grip) Needle holder (Duro grip) Needle holder (Duro grip) Needle holder De Bakey Needle holder De Bakey Needle holder De Bakey Needle holder De Bakey Needle holder De Bakey Hagar Mayo seedy needle holder Wangansteen needle holder Hegar mayo needle holder Hegar mayo needle holder Hegar mayo needle holder Hegar needle holder Hegar needle holder Ryder needle holder Ryder needle holder Ryder needle holder Ryder needle holder Ryder needle holder Crile wood needle holder Mayo hegar needle holder Mayo hegar needle holder Mayo hegar needle holder Jameson needle holder Massan needle holder Castroviejo needle holder Castroviejo needle holder Barraquer needle holder straight Barraquer needle holder straight with catch Reill needle holder straight Reill needle holder with catch straight Finochietto needle holder Skin straightening plate Mouth gag set Turner warvick ring retractor with blade Turner warvick needle Cat.no. BF 437 BM-25 BM-26 BM-27 BM-35 BM-36 BM-37 BM-51 BM-52 BM-22 BM-42 BM-65 BM-66 BM 67 BM-77 BM-78 BM54 Bm 55 BM 56 BM 57 BM-47 BM 220 BM 230 BM 236 BM 237 BM 270 BM 275 BM 563 BM 564 FD 230 FD 232 FD 240 FD 242 BM 277 BA 709 Specification 100mm 4" 180mm 7" 200mm 8" 230mm 9" 180mm, 7" 230mm 9" 250mm 10" 195mm 7 ¾" 245mm 9 ¾" 200mm 8" 275mm 10 ¾" 150MM 6" 180mm 7" 200mm 8" 200mm 8" 240mm 9 ½" 130mm 5 1/8" 150MM 6" 180mm 7" 200 mm 8" 265mm 10 ½" 160mm 6 ¼" 150mm 6" 180mm 7" 200 mm 8" 235mm 9/14" 265mm 10 ½" 140mm 5 ½" 140mm 5 ½" 120mm 4 ¾" 120mm 4 ¾" 150mm 6" 150mm 6" 260mm 10 ¼" Qty. 12 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 4

39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70

Baby mosquito (Hartmann) straight Baby mosquito (Hartmann) curved Micro halstead mosquito straight Micro halstead mosquito curved Halstead mosquito straight Halstead mosquito curved Rankin haemostatic forceps straight Rankin haemostatic forceps curved Pocket instrument set Breiskey Narvatil retractor With hook Langenbeck green retractor Longen beck green retractor Methew retractor Dever retractor Dever retractor Dever retractor Dever retractor Dever retractor Dever retractor Dever retractor Dever retractor Kocher retractor Kocher retractor Kocher retractor Langen beck retractor Retractor trachea Retractor trachea Retractor trachea Retractor trachea Vein retractor cushing Vein retractor cushing Vein retractor cushing kocher

BH 104 BH 105 BH 108 BH 109 BH 110 BH-111 BH 164 BH 165 BH 807 BT 177 BT 178 BT 40 BT 600 BT 601 BT 606 BT 602 BT 603 BT 604 BT 605 BT 608 BT 461 BT 462 BT 463 BT 326 BT 111 BT 112 BT 113 BT 114 BT 184 BT 185 BT 190

95mm 3 ¾" 95mm 3 ¾" 125mm 5" 125mm 5" 125mm 5" 125mm 5" 160mm 6 ¼" 160mm 6 ¼" 03 set 06 set 160mm, 6 1/3" 16x6mm 160mm 6 ¼" 25x6mm 190mm 7 ½" 180mm 7" 20mm 215mm 8 ½" 22mm 230mm 8 ½" 22mm 300mm 12" 25mm 330mm 131/8" 5mm 300mm 12"38mm 300mm 12" 50mm 300 mm 12"75 mm 240mm 9 ½" 80x42mm 240mm 9 ½" 80x50mm 240mm 9 ½" 80x60mm 210mm 8 ¼" 42x13mm 160mm 6 ¼" 160mm 6 ¼" 160mm 6 ¼" 160mm 6 ¼" 200mm 8"11x14mm 200mm,8" 13x17mm 180 mm 7"

24 24 24 24 24 24 24 24 2 2 2 4 4 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95

Dissecting forceps rumel Skin retractor kilner Skin retractor Gillies Skin retractor fisch Guyon catheter guide Kocher intestinal clamp Fergusson suction instrument (olive) Fergusson suction instrument (olive) Fergusson suction instrument (olive) Fergusson suction instrument (olive) Fergusson suction instrument (olive) Babcock Babcock Debakey tissue grasping forceps De Bakey tissue grasping forceps Baby allis Baby allis Allis Thomas Rib spreader finochietto Burford Self retaining retractor weitlaner Self retaining retractor weitlaner Abdominal retractor Oettingen Abdominal retractor robin masse Kidney basin medium size Bowel medium size

BJ 45 OL 610 OL 611 FD 376 EF 492 EA 156 GF 927 GF 928 GF 929 GF 930 GF 931 EA30 EA 32 FB 400 FB 402 EA 10 EA 12 EA 20 FB 805 BV 71 BV 72 BV 537 BV 541

10x10mm 230mm 9" 150mm 6" 180mm 7" 180mm 7" 365mm 14 ½" 280mm 11" 2.0mm 2.5mm 3.0mm 4.0mm 5.0mm 155mm 6" 215mm 8 ½" 150mm 6" 200mm 8" 150mm 6" 120mm 4 ¾" 200 mm 8" 130mm, 5 1/8" 165mm 6 ½" Complete set Complete set

2 2 2 2 2 2 2 2 2 2 2 24 24 4 4 12 12 24 2 2 2 2 2 4 4 Annex-I

List of microsurgical instruments for Urology Operating Theatre S.no. 1 2 3 4 5 Name of instrument Needle holder (Duro grip) Needle holder (Duro grip) Needle holder (Duro grip) Needle holder De Bakey Needle holder De Bakey Specification 180mm 7" 2 200mm 8" 2 230mm 9" 2 180mm, 7" 2 230mm 9" 2 Qty.

6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 34 35 36 37 38 39 40 41 42 43 44 45 46 47

Needle holder De Bakey Needle holder De Bakey Needle holder De Bakey Hagar Mayo seedy needle holder Wangansteen needle holder Hegar mayo needle holder Ryder needle holder Ryder needle holder Crile wood needle holder Mayo hegar needle holder Mayo hegar needle holder Massan needle holder Castroviejo needle holder Castroviejo needle holder Finochietto needle holder Skin straightening plate Mouth gag set Turner warvick ring retractor with blade Turner warvick needle Baby mosquito (Hartmann) straight Baby mosquito (Hartmann) curved Micro halstead mosquito straight Micro halstead mosquito curved Halstead mosquito straight Halstead mosquito curved Rankin haemostatic forceps straight Rankin haemostatic forceps curved Pocket instrument set Breiskey Narvatil retractor With hook Retractor trachea Vein retractor cushing Vein retractor cushing Dissecting forceps rumel Skin retractor kilner Skin retractor Gillies Fergusson suction instrument (olive) Fergusson suction instrument (olive) Babcock Babcock Debakey tissue grasping forceps De Bakey tissue grasping forceps Baby allis

250mm 10" 195mm 7 ¾" 245mm 9 ¾" 200mm 8" 275mm 10 ¾" 180mm 7" 130mm 5 1/8" 150MM 6" 160mm 6 ¼" 150mm 6" 180mm 7" 265mm 10 ½" 140mm 5 ½" 140mm 5 ½" 260mm 10 ¼"

95mm 3 ¾" 95mm 3 ¾" 125mm 5" 125mm 5" 125mm 5" 125mm 5" 160mm 6 ¼" 160mm 6 ¼"

160mm 6 ¼" 200mm 8"11x14mm 200mm,8"13x17mm 230mm 9" 150mm 6" 180mm 7" 2.0mm 2.5mm 155mm 6" 215mm 8 ½" 150mm 6" 200mm 8" 150mm 6"

2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 4 36 36 36 36 36 36 24 24 2 3 set 6 set 2 2 2 2 2 2 2 2 36 24 4 4 12

48 49 48 49 50 51 52 53 54 55 56 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86

Baby allis Magill forceps Magill forceps Magill forceps Scalpel handle Scalpel handle Scalpel handle Steven's scissor dissecting Steven's dissecting scissor Reynolds ­ Jameson vessel scissor Joseph straight scissor Metzenbaum dissecting scissor Metzen baum dissecting scissor Baby ­ Metzen baum straight scissor Baby- Metzen baum curved very delicate Toennis ­ Adson curved very delicate scissor Toennis-Adson curved delicate scissor Toennis straight very delicate scissor Toennis straight delicate scissor Toennis curved scissor Mayo-stille straight scissor Mayo-stille curved scissor Mayo-Harrington straight scissor Mayo-Harrington curved scissor Mayo-straight scissors Mayo curved scissor Metzenbaum straight scissor Metzen baum straight scissor Metzen baum curved scissor Metzen baum curved scissor Pott's Demartel vessel scissor 25 degree Pott's Demartel vessel scissor 45 degree Pott's Demartel vessel scissor 60 degree Baby Metzen baum curved scissor Metzen baum straight scissor Metzen baum straight scissor Metzen baum curved scissor Metzen baum curved scissor Metzen baum curved scissor

120mm 4 ¾" 15mm, 6" 200mm, 8" 250mm, 10" No.3 No. 4 No. 7 115mm, 4 ½" 115mm, 4 ½" 114mm, 5 ½ " 150mm, 6" 145mm, 5 ¾" 145mm, 5 ¾" 145mm, 5 ¾" 145mm 5 ¾ " 175mm, 7" 175mm, 7" 180mm, 7" 180mm, 7" 180mm, 7" 170mm, 6 ¾ " 170mm 6 ¾ " 225mm, 9" 225mm, 9" 195mm, 7 ¾ " 210mm,8¼ " 180mm, 7" 200mm, 8" 180mm, 7" 200mm, 8" 190mm, 7 ½ " 190mm , 7 ½ " 190mm, 7 ½ " 145mm, 5¾" 180mm 7" 200mm, 8" 230mm, 8 " 180mm 7" 200mm 8"

12 4 4 4 4 4 4 4 4 4 2 2 4 4 4 2 2 2 2 2 4 4 4 4 4 4 4 4 4 4 4 4 4 4 2 2 2 2 2

87

Metzen baum curved scissor

230mm, 9"

2

88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 129 130 131 132

Nelson-Metzen baum curved scissor Nelson-Metzen baum curved scissor Nelson Metzen baum curved scissor Nelson Metzen baum curved scissor Metzen baum curved scissor Micro scissor straight Micro scissor curved Micro scissor straight Micro scissor curved Micro scissor straight Micro scissor curved Dissecting forceps cushing Dissecting forceps cushing Dissecting forceps cushing Micro Adson dissecting forceps Adson dissecting forceps Tissue forceps very delicate Tissue forceps Waugh Tissue forceps borrey Lung dissecting forceps Micro Adson tissue forceps Micro Adson tissue forceps Micro Adson tissue forceps Tissue forceps standard Gillies tissue forceps Tissue forceps Non traumatic dissecting forceps Adson Non traumatic dissecting forceps Adson Non traumatic dissecting forceps Adson Non traumatic dissecting forceps cushing Non traumatic dissecting forceps Standard Wangesten dissecting forceps Wangesten dissecting forceps Micro forceps jeweler type Reynold's dissecting scissor Reynold's dissecting scissor Littler dissecting scissors Nelson metzen baum dissecting scissor Nelson metzen baum dissecting scissor Nelson metzen baum dissecting scissor Nelson Metzen baum dissecting scissor Weller dissecting scissor Foerster Ballenger sponge holding forceps Foerster Ballenger sponge holding forceps Non traumatic vascular tissue forceps De Bakey

250mm, 10" 280mm, 11" 230mm, 9" 230mm 9" 230mm 9" 105mm, 41/8" 105mm 4 1/8" 80mm, 3 1/8" 80mm 3 1/8" 120mm 4 3/8" 120mm 4 3/8" 180mm 7" 200mm 8" 250mm 10" 120mm 4 ¾ " 120mm 4 ¾ " 100mm 4" 180mm 7" 180mm, 7" 230mm 9" 120mm 4 ¾" 120mm4 ¾" 120mm 4 ¾" 180mm 7" 160mm 6 ¼" 169mm 6 ¼" 120mm 4 ¾" 145mm 5 ¾" 156mm 7" 180mm 7" 145mm 5 ¾" 230mm 9" 255mm 10" 110mm 4 3/8" 145mm 5 ¾" 180mm 7" 120mm 4 ¾" 230mm 9" 250mm 10" 230 mm 9" 250mm 10" 280mm 11" 240mm 9 ½" 240mm 9 ½" 150mm 6"

2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

133 134 135 136 137 138 139 140 141 142 143 144 145 146 147 148 150 151 152 153 154 155 156 157 158 159 160 161 162 163 164 165 166 167 168 169 170 171 172 173 174 175

Non traumatic vascular tissue forceps De Bakey Bull dog clamp Dieffen bech Bull dog clamp dieffen bech Bull dog clamp De Bakey Hess Bull dog clamp De Bakey Hess Bull dog clamps Bull dog clamps Bull dog clamps Bull dog clamp Bull dog clamp De Bakey Bull dog clamp De Bakey Bull dog clamps De Bakey Bull dog clamps De Bakey Bull dog clamps De Bakey Bull dog clamps De Bakey Bull dog clamp De Bakey Bull dog clamp De Bakey Bull dog clamp De Bakey Bull dog clamp Glover Bull dog clamp Glover Bull dog clamp Glover Bull dog clamp Glover Bull dog clamp Glover Bull dog clamp Glover Non traumatic vessel clamp Leland Jones Non traumatic vessel clamp De Bakey Non traumatic vessel clamp De Bakey Non traumatic vessel clamp Non traumatic vessel clamp Non traumatic vessel clamp Non traumatic vessel clamp Non traumatic vessel clamp Non traumatic vessel clamp Non traumatic vessel clamp Baby satinsky vessel clamp Satinsky vessel clamp Satinsky vessel clamp Satinsky vessel clamp Vessel scissor De Bakey Vessel scissor De Bakey Vessel scissor Diethrich Hegemann Vessel scissor Diethrich Hegemann

200mm 8" 35mm 1 3/8" 50mm 2" 20mm 30mm 35mm 1 3/8" 35mm 1 3/8" 45mm 1 ¾" 45mm 1 ¾" 50mm 2" 50mm 2" 70mm 2 ¾" 85mm 3 3/8" 105mm 4 1/8" 105mm 4 1/8" 120mm 4 ¾" 120mm 4 ¾" 27 mm 1 1/8" 65mm 2 ½" 65mm 2 ½" 65mm 3 ¾" 95mm 3 ¾" 115mm 4 ½" 115mm 4 ½" 190mm 7 ½" 190 mm 7 ½" 230 mm 9" 195mm 7 ¾", Jaws 35mm 260mm 10 ¼",Jaws 45mm

2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

265 mm 10½",Jaws 55mm 275mm 10 ¾",Jaws 60mm 280mm 11",Jaws 75mm 200 mm 8",65mm Jaws 265 mm 10½", Jaws 70mm 150mm 6" 265mm 10 ½" 265mm 10 ½" 265mm 10 ½" 165mm 6 ½" 165mm 6 ½" 175mm 7" 175mm 7"

176 Vessel scissor Diethrich Megemann 175mm 7" 177 Micro Adson Dissecting & legasure forceps 140mm 5 ½" 178 Baby mixter dissecting forceps 140mm 5 1/2m 179 Baby Adson dissecting forceps 140mm, 5 ½" 180 Baby mixter dissecting forceps 180mm 7" 181 Baby adson Dissecting forceps 180mm 7" 182 Lahey (sweet) dissecting forceps 190mm 7 ½" 183 Lahey dissecting forceps 230mm 9" 184 Mixter dissecting forceps 220mm 8 ¾" 185 Mixter dissecting forceps 250mm 10" 186 Lawrence dissecting forceps 270mm 10 ¾" 187 Castroviejo needle holder 140mm 5½" 188 Converse needle holder 125mm 5" 189 Masing needle holder 130mm 5 1'2" 190 Halsey needle holder 130mm 5 ½" 191 Baby crilewood needle holder 192 Schink skin grafting knives 300 12" 193 Kersting sigmoid forceps 300mm 12" 194 Millin's retracrtor complete set 195 02 pair of blades 196 Scott Perineal retractor 197 Sponge holder Forceps Ring tip forceps 9 inch vascular (Titanium) Coman's Scissors Pott's right angle scissors. Trocar Paediatric 3.5mm trocar for laparoscopic surgery Forceps Marry land forceps 3.5mm L=300mm Needle holder 3.5mm, L=300mm Scissor 3.5mm, L=300mm Hook Dissecting hook 3.5mm Bipolar cautery 3.5mm, L=300mm Grasping forceps Grasping forceps 3.5mm, L=300mm

2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 1 2 1 2 12

Technical Specifications of OPMI surgical microscope camera unit Microscope body 5 step apochromatic magnification changer, magnification factors 0.4/0.6/1.0/1.6/2.5, Integrated swing-in red free green and orange filters. Widefield strainght/180 degree inclined, focal length f=170mm interpupillary distance adjustable from 55mm to 75mm. High- eyepoint widefield push-in eyepieces 10x or 12.5x with magnetic locks.

Binocular tube Eyepieces

Objective lenses

Interchangeable objective lenses f=200/250/300/350/400 mm , Interchangeable objective lenses f=200/250/300mm with focusing devices and sterilizable caps, fine focusing range: 13mm . T-shaped grip, handgrips (back or front) with sterilizable caps, comfortable handgrips (back with removable sterilizable handgrips. 120 degree coupling, swivel coupling with brake, coupling K 90/60. Coaxial fibre optic halogen illumination system 12V 100 W with fast action lamp changer. S100 floor stand/ceiling mount/wall mount and table mount. Integrated video camera : 1CCD, 752 (H) x 582 (V) pixels (PAL), 768 (H) x 494 (V) pixels (NTSC), S/N ratio 50 dB, horizontal resolution 480 lines, Y/C (S-Video) and VBS (composite) signal outputs, foot panel connector for image freeze. Endoport: Optical port for attaching all standard endoscopic cameras with high-speed shutters. Beam splitters for co-observation and documentation devices. Co-observation devices (mono/stereo). Video objective lenses f=50/60/85/105 mm with C-mount Video camera 1CCD and 3 CCD. C-mount adapter for Nikon Coolpix digital cameras. Asepsis/sterilizable caps for controls.

Handgrips

Couplings Illumination system Suspension system Accessories

SPECIFICATIONS PATHOLOGY DEPARTMENT Name of item with full specifications PATHOLOGY DEPARTMENT (Hematology) Slide Stainer (Semi automated/automated) will all standard accessories and given specifications: - Semi-automated/automated bench top compact slide stainer

- Microprocessor controlled, menu driven and programmable - Continuous load and upload of slide racks/ basket - Over 10 racks/baskets each with 20-25 slide capacity. - 5-10 reagent stations holding upto 100-500 ml of reagent - 230 + 20V, 50 Hz electrical specification (Availability of slide drying facility will be an additional feature)

SPECIFICATION FOR PORTABLE COLD LIGHT Portable cold light for 4 focus portable OT light with accessories OT (to be purchased from any reputed Indian firm) Xenon cold light Xenon video cold light foundation , 300 Watt (NOVE) with cable.

SPECIFICATION FOR FLEXIBLE CYSTOSCOPE SET S.No Nomenclature with full specification . 1 Flexible Cystoscope Set Specifications:-Instrument Channel. -Flexible biopsy forceps. -Flexible foreign body forceps. Note: It should be purchased at par the specifications of Karl Stroz or Richard Wolf's instruments).

SPECIFICATION FOR FLEXIBLE URETEROSCOPE SET S.No Nomenclature with full specification . 1 Flexible ureteroscope Set Specifications:-Instrument Channel. -Stone holding flexible forceps. -Flexible biopsy forceps. Note: It should be purchased at par the specifications of Karl Stroz or Richard Wolf's instruments).

Specifications for Electrophoresis system Electrophoresis system with Power supply: a) Mini horizontal Gel unit with removable Gel casting tray (includes UV tray, 2x1.0mm 8-well combs). b) Single plate mini gel, without casting base 10x10 cm (includes integral gel running module, 2 plane glass plates with 1mm thick bonded spacers, 2 notched glass plates & one 12 well, 1mm thick comb). c) Power station / Power supply for electrophoresis: Constant voltage or constant current with automatic crossover, Timer: 0 to 999 minutes, Display: 3 digit LED, Output terminals: 4, parallel, Dimensions: 3x22x7.4 cm, Power station 300 plus, 10-300 VDC, 4-500 mA. Specifications for S.S. Container Cost per unit Total cost Sl Name of item with fill Qty. (in (approx.) in (Approx.) No specification figure and words) 1. S.S. Container Size: 60X30X30Cms. 35 Nos. 60X30X15Cms. 10 Nos. 30X30X30Cms. 45 Nos. 30X30X15Cms. 05 Nos. 2. Perforated Tray S.S. Size: 48X26Cms. 30 Nos. 24X24Cms. 35 Nos. 3. Textile Filters 2000 Nos. 54.5Cms.X24Cms. Note:1. The Containers should have lid with internal filter and side vents. 2. Base of Containers should be perforated with filter & filter sheet (Textile filters). 3. All parts of containers may be made up of chromium Nickel sheet. 4. Each container should be supplied with textile filter. 5. The containers were purchase/ Procured from M/s Ceatec Medizintechnik/M/s Martin, Germany. 6. This is remaining demand of previous P ­ 2 of year 2003. These are needed for Packing & Sterilization of Operation theatre Linen / Instrument

Specification of Table Top Autoclave (Immunology) Direct steam operated with built in steam generator and with proper air breaker / release system. The complete chamber should be of SS 304 and door manually lockable. The sterilizer should be provided with lid and tray. It should have door within electrical / mechanical safety. Maximum wattage of not more than 2 KW. Dual operating temperature of 121ºC and 134ºC; all the components other than drain cock should be of SS.

Specification of Autoclave vertical (Microbiology) Chamber material: Stainless steel Capacity- 50 L Sterilization temperature: 1050C ­ 1310C (2210-267.80F) Temperature gauge range: Digital display 800C-1410C Safety valve releasing pressure: 177kPa (25psig) Pressure gauge range: 0-0.3Mpa/0-45psi Timer setting range: 1-180 minutes Exhaust tank: 3 L polyethylene tank Safety devices: Pressure safety valve anti dry scorch thermo limiter Door switch handle switch current fuse Three years warranty with AMC (comprehensive). · · · · · · · · ·

Specification for Electrophoresis System 1. 96 well electrophoresis system for HLA typing: Horizontal Gel unit with removable gel casting tray with a comb having 96 wells.

Specification for Anaesthesia Workstation

Integrated Anaesthesia workstation for advanced surgical setup. Comprising of anaesthesia delivery system, Compact rebreathing system, Two agent specific vaporizers and ICU quality ventilator for adult and children with advance modes like PCV, PSV & SIMV etc.

Ergonomic design with coloured graphic user interface, and data port with Oxygen, Gas flow and ventilator information available in electronic form. Anaesthesia machine : Unit should have primary connection for Central gas supply with pressure gauges indicating inlet line pressure of all three gases i.e. Air, Oxygen and Nitrous Oxide. As a backup machine should also have provision for connecting oxygen and nitrus oxide pin index cylinders. Machine should have Electronic fresh Gas control / monitoring of flows. Audio - visual alarm for failure of Oxygen. Control of minimum 21% Oxygen in fresh gas upto Folw > 1L/min. and atleast 250 ml of Oxygen concentration for minimal flow application (fresh gas flow < 1L/min). No basal flow. Integrated Oxygen flush with self returning valve. Machine should also have an independent fresh gas outlet for connection to Bain's or Magill circuit. Fresh gas flow setting from 50ml/min to 12 Ltr/min Breathing system : Compact breathing system suitable for minimal flow anaesthesia, with least patient circuit volume including absorber etc. approx. 3 L (excluding bag) for fast response to change in fresh gas composition. Fresh gas decoupled / compensated breathing system for adult and children with possibilities to mount the breathing system on left or right side. APL valve with direct setting of release pressure.

Integrated Ventilator : Ventilator suitable for adult and children without changing of bellow. Automatic breathing circuit Compliance correction. Light weight bellow should not offer any constant PEEP. Spont. Breathing Manual Ventilation IPPV with Platu adjustment from 0 to 50% of Ti

PLV with decelarating flow Pressure controlled ventilation PCV Pressure Support & SMIV. High peak inspiratory flow upto 70 LPM, Tidal volume adjustment range 20ml to 1400 ml PEEP from 0 to 20 mbar Electronically adjustable Resp frequency from 6 to 60 per min. I:E from 1:4 to 4 :1

Single step changeover of mode. Ventilator should have standard PVO (Airway pressures, Volumes and Oxygen) monitoring. Easy to start with auto set alarms and a central color graphic display for settings and monitored values Vaporizers : Temperature / pressure compensated and flow independent Halothane and Isoflurane vaporizer. Vaporizer should have transport lock to provide hermetic sealing of agent chamber during transport & storage. The vaporiser design should be maintenance free. Should not require periodic overhaul etc. As per manufacturer recommendation. Anaesthesia Monitor : Atleast 6-8 channel monitor with minimum 12" to15" color (Active matrix display) screen. Suitable for adult, pead. & neonatal patients. Should have standard parameters ­ ECG ­ 5 Lead, SPO2, NIBP, 2 X Temperature, Respiration, 2 X IBP, Respiratory gas module and capnography. Graphic & Tabular trends for 24 Hrs Arrhythmia detection & Alarm 3 Lead ST Segment analysis

Scope of Supply : Three gas Anaesthesia workstation Trolley with drawers and locking Full length side GCX rails for mounting of accessories Yokes ­ 2 X Oxygen & 1 X N2O, besides central supply connections Integrated ventilator and semi-closed breathing system Two agent specific vaporisers for Halothane and Isoflourane Reusable adult and Peadiatric patient tubings set ­ 5 adult and 5 pediatric with each machine

Central gas supply hoses Accessories for adult & paediatric including all sensors ­ ECG 5 lead set - 2 NIBP cuff ­ 4 sizes ­ 5 set with each Machine SpO2 sensor ­ Adult & Pead ­ 3 set with each Machine Temp sensor ­ Skin & Rectal ­ 5 set with each Machine I.B.P. should be without transducer Suitable mount for mounting the monitor on Anaesthesia machine A Gases module for Anaesthetic agents and N2O (With auto agent identification) including capnography. BIS module Warranty : ­ 5 Years.

Specifications of Cryostat

Compact open-top-cryostat, with heated removable sliding window and cryochamber illumination; counterbalanced handwheel, lockable in the top position; cross roller guides. Cryochamber temperature selectable between 0°C and -30°C. Actively cooled freezing shelf for up to 10 specimen discs; maximum freezing shelf temperature -45°C, depending on the selected cryochamber temperature (20°C to -30°C). 8-minute evaporator defrost cycle which can be activated manually at any time and / or programmed. Slot cover. Single screw locking system for direct fitting of specimen discs; Precision X/Y/Z 8° specimen orientation; removable microtome from the cryochamber; Section thickness selectable from 160 µm adjustable in multiple steps; Total horizontal specimen feed : approx.25 mm; 2 Motorized coarse feed speeds : 0.3 mm/s and 0.65 mm/s FOR PATHOLOGY Specifications of Height Adjustable Cryostat The equipment offer must be of latest model Open top, heated sliding window, corrosion proof stainless steel cryochamber with good illumination. Microprocessor based touch key control panel with LCD display for all functions including of microtome. Separate control for specimen head and knife edge temp. up to (-) 30 degree C or less. Temperature control of cryochamber should be at least (-) 30 degree or less. Preference will be given to models with dual compressor. Facility for integrated quick specimen freezing up to (-) 60 degree C. Space for other specimen rack. Removable section waste tray. Automatic programmable defrosting with enclosed drainage system of defrost fluid. Holders / carriers for holding and using minimum 16 cm "C" type knife, disposable high and low profile blades. Facility for magnetic fixation of knife is preferable. Specimen holder can hold specimen size up to 70x50 mm. Facility for both 360 degree rotation as well as movement in xy axis Fully motorized microtome: movement control by manual as well as by foot switch. Facility for hand wheel brake. Section thickness cutting 1-60 micrometer, Specimen retraction around 50 micron.

Cryochamber should be of standard size (approximately 50x40 cm and able to handle at least 10 blocks at a time). Power requirement: 220/240 volts, 50 Hz AC supply. Standard accessories to be provided with the main equipment should be specified. Servo stabilizer of appropriate rating should be provided. Detailed product catalogue, operative and service manuals must be provided with the equipment. Rights for minor specification changes / technical changes reserved. Comprehensive warranty for three years and also quote for comprehensive and non-comprehensive warranty from 4th year to 10th year. Apart from standard accessories, rates for following accessories be also quoted. Cryostat oil - 5x100 ml Brushes - 20 Anti roll plates - 02 Freezing medium - 30x100 ml High profile blades - 10 packet (pack size of 50 blades) Low profile disposable blades - 5 packets (pack size of 50 blades) Type "C" knife 16 cm - 05 Cryochamber illumination fluorescent lamp - 02 Specimen holders of different sizes 20 The details and rates for non-motorized microtome model of cryostat with aforementioned specification should be quoted separately. SPECIFICATIONS Digital High resolution video camera Digital High resolution video camera for Microscope Resolution 2588 x 1960 Specification 3.4x3.4um 3.5fps at 2560 x 1960 (30-bit) High speed preview and capture options in software universal microscopy imaging in biomedical and materials applications SPECIFICATION FOR DIRECT DIGITAL RADIOGRAPHY SYSTEM Direct Digital Imaging System for General Radiography should be Robust and Manoeurable in order to undertake Wide Range of Examination Required. The design of the system should feature a wide range of Movements and Incorporate ergonomic considerations. A Single Active Matrix Flat Panel Detector, which can be, moved into different positions for all radiographic projections viz. Vertical, Table and Extremities. The Detector may be moved

automatically or manually in different positions. There should a be free movement of the detector between the x-ray table and vertical stand. Motorized table will be accepted. Mention the Year of Introduction of the Quoted Model in the International market and the Quoted Model should be latest state of art. Give details of all pre-installation requirement (The firm are free to visit the installation Site and also to see dry laser camera before Quoting). A) Detailed Specification of X-ray Flat Panel Detector (Quote the latest model of Flat Panel Detector)

Note: Specifications to be supported by compliance statement, technical data sheet of detector and any other relevant material. The buyer reserves the right to call for any additional information from the seller at the time of technical evaluation. 1. 2. the firm. 3. 4. process. 5. Scintillator material used for flat panel detector: Thallium Doped Cesium Iodide (CsI: TI). 6. Semi Conductor material (Photodiode): Hydro generated Amorphous Silicon (a:Sl:H). 7. Charge Read Out: Thin Film Transistor (Array) (TFT Array). 8. Detector Size: 41 cm x 41 cm2 or more. 9. Array size: 2000x2000 pixel or more. 10. Pixel pitch: 0.2 mm or less. 11. Image depth: 14 bits or more. 12. Wide Dynamic Range of Flat Detector. 13. Detector quantum efficiency (DQE): Specify the DQE at what spatial frequency. 14. Mention the weight of the detector. 15. Cable length: Cable length to be sufficient for movement of single active matrix flat panel detector in different position for all radiographic applications (If applicable). 16. Mention the Detector assembly size (Hx Wx Dcm). 17. Mention any temperature to be maintained for optimum performance of detector. 18. Mention any cooling required for the detector. 19. Mention whether any calibration is required or not. If needed how frequent and by whom. 20. Mention the Estimated life of the detector. Assembling: Tiled or Untiled Panel. Active matrix flat panel detector: Based on Indirect conversion Use of Active Matrix Flat Panel Imager: (Radiography). Name of the Detector Model and manufacturer: To be provided by

21. available). 22. 23.

Mention any other version of Indirect Flat Panel Detector (model System warm up time should be minimum. Important: Option should be available for the system to be independently used for film screen and Computed Radiography (Imaging plates of different size).

B) Specification of Acquisition / Review workstation: High resolution Medical Grade Monitor for Digital Radiography. Monochrome LCD monitor with Protect panel from dust and scratches. Manufacturers Name and Model: Viewing Angles (H, V): 1700 or more Size of Monitor (diagonal) 21" or more Mouse control display or Touch screen (Optional) Mention all the standard accessories to be supplied with the Monitor Luminance (cd/m2) and contrast ratio should be very high: Specify Mention the specification for Compatible graphic controller board for the Medical Grade Display 10. Three Mega pixel monochrome LCD Monitor (Native Resolution 2048x1536) in Portrait and Landscape Mode 11. Mention the number of LCD monitors needed for Direct Digital Radiography [Note: If there are 2 monitors one monitor will be High Resolution Medical Grade and the other Monitor which will be for text, protocols, X-ray Controller application need not be a Medical Grade Monitor {mention the size, resolution and other details}] 12. Calibration Support and Quality Control Management according to DICOM PART 14 standard or any other equivalent or higher version 13. Hard Disk Storage: 70 GB or more Storage of 3000 or more images. 14. Fast Acquisition, Image Processing and Display: Mention the time 15. The system should have the facility of integrating with Dry Laser Camera of Fuji make 7000dry pix and other models of camera for printing (DICOM Complaint) C) X-ray Table Specification: 1. Table should be motorized with floating table top (having lateral and longitudinal movements) 2. Elevation and Downward Movement of Table should be Motor Driven 3. Foot pedals provisions 4. Maximum Patient Weight: 200 kg or more 5. Table Top Length: 200 cm or more 6. Table Top Height: Minimum 50 cm or above, Maximum 80 cm or above 7. Provision for Inherent filtration with Aluminum equivalent of table top 8. Movement of table top or detectors for maximum patient coverage: specify 9. Anti scatter radiation grid: Focus grid Minimum 100 cm and mention the Grid lines and Grid ratio 10. Anti collision device facility 1. 2. 3. 4. 5. 6. 7. 8. 9.

11. Servo controlled movement to X-ray tube along with the Detector (Simultaneous Movements) D. Vertical stand (Specification) 1. Vertical movement: Manual and Motorized with foot switch facility 2. The vertical movement to be servo coupled to the movement of the x-ray tube (Simultaneous movements) 3. Standard removable Focus Grid (Minimum Two) Mention the number of lines/cm and Grid ratio 4. Mention if tilting detector facility available 5. Box to be provide for keeping Removable grid 6. Height of detector panel (center) from the floor should be minimum 25 cm 7. Maximum height from the floor to the center of detector should not be than 175cm E. Ceiling Mounted X-ray Tube: a. b. c. d. X-ray Tube suspended on a column The movement of X-ray Tube should be motorized and movement in all possible directions: Specify the Travel Range and Angulations in Degrees It should have capability of Manual Override Provision for Patient Side Control Panel

F. X-Ray Generator Specification: a. Invertors Type Constant Potential High Voltage Generator (High Frequency X-Ray Generator) b. Power : 80 KW c. 800 mA at 100 kV according to IEC standard d. kVp : 40 to 150 e. Automatic Exposure control G. X-ray Tube (Specification) a. b. c. d. e. f. g. Anode Heat Storage Capacity: 300 KHU or more Mention the name of the X-ray tube Specify the Heat Storage Capacity of the whole tube Specify the cooling rate of the Anode and Tube Inherent filtration with aluminum Automatic Collimator with Manual Override Facility Dual focus 0.6 mm and 1.25 mm or less.

H. Accessories 1. Compression belt (Pediatric and adult) [ 1 or 2 each] 2. Patient hand grip

3. 4. 5. 6. 7. 8. 9.

Foot switch for adjusting the height of the stand or the table Lateral cassette holder Patient support bar for vertical stand Lead Glass 80 cm x 100 cm Table top locking facility Specify any Stabilizer required for the system UPS of appropriate rating (with Half hour back up) for the acquisition/Review workstation and a 2nd monitor (if supplied with the system) of reputed brand

I. Advanced Clinical Application Facility: 1. 2. 3. 4. Image Stitching / Image Pasting for Complete spinal column, Extra Long Leg Image etc (Quote as standard) Specify the Maximum number of images which can be pasted together for different organs (quote as standard) Dual Energy subtraction facility (quote as standard) Mention about computed aided detection (CAD) available with the system. Provide all the details and technical aspects (Quote as Optional)

J. Pre Processing Facility should be available like 1. 2. 3. 4. 5. 6. Collimation Recognition Rejection of Unattenuated beam Histogram Processing Application of look up table Frequency dependent edge and contrast enhancement Any other facility available- Please mention

K. Post Processing Facility should be available like: 1. 2. 3. 4. 5. 6. 7. 8. Addition of Anatomical Markers Demographic Correction Image Annotation Window & level Adjustment Electronic Collimation Magnification Application of different LUT's Any other facility available (please mention)

L. Dry Laser Imager Consisting of · Throughput 50 films (14"x17")/hour or more · · Spatial resolution of more than 500 DPI/PPI Print contrast resolution 12 bits/pixel

· · ·

Multiple image formats (specify) Minimum 2 online sizes Different film sizes possible. 14" x 17" 11" x 14" 14" x 14" 10" x 12" 8" x 10"

· · ·

Necessary accessories including magazines etc. Compatibility with different brand of films. DICOM ready system.

M. Networking (Specifications): 1. Patient Data Entry: HIS / RIS, Keyboard & mouse or bar code reader 2. The System should be capable to connect to PACS through RIS/HIS 3. DICOM 3.0 or higher version with a. Modality push b. Modality work list c. Print Service d. PACS Push e. Performed Procedure Step f. PACS confirm storage g. Modality Pull h. PACS full i. Modality verification j. Storage of image on CD/DVD k. Ability to send, receive, acknowledge, print and record on CD/DVD l. Connectivity to PC m. Mention any other service N. Other terms and conditions: 1. Some specification which are not quantified, the buyer reserves the right to evaluate the specification based on the details provided by the firm 2. The equipment should be under comprehensive warranty for five years from the date of successful installation and handing over. 3. Please quote Comprehensive Maintenance Contract (Including X-ray Tube and detector) for five years after successful completion of warranty

4. The firm will provide applications engineer for the first two weeks to assist functioning on the new equipment. The firm will also render on-site training to doctors and technicians. 5. All software Up-gradations will be provided free of cost to the Institute as and when available. 6. Operating manual, Service Manual along with Schematic Diagram to be provided. 7. There will be an agreement between the Buyer and Seller for Comprehensive Maintenance Contract at the time of finalization of purchase of equipment. Up time Guarantee : 95% of 365 working days per year. Downtime exceeding 5% (five) of 365 working days per year will be penalized at the rate as per final agreement. 8. Only Principals or their authorized principal Indian agents should participate in the tender. 9. Image quality assurance as per latest IPEM standard. 10. All the civil, electrical alterations/fixation pertaining to the installation of the machine will be the responsibility of the firm. The Institute however shall provide the room for installation and one electrical point of power supply, air-conditioning system. Tender Specification for Dual Head SPECT (CT) Gamma Camera General: A latest technology Dual Headed variable angle SPECT (CT) Gamma Camera system capable of performing all planar, SPECT, WB SPECT and Whole Body imaging applications with CT based transmission attenuation correction and functional anatomical image fusion. · · · · · · · · Dual Headed variable angle SPECT (CT) system with minimum of four or more slice CT. SPECT & Whole body imaging capabilities. CT based attenuation correction. Diagnostic CT image acquisition and fusion with nuclear medicine image. Image fusion software & hardware. Features of diagnostic CT should include other conventional features of required hardware and software. Capable of acquiring multi slices SPECT/CT scan of full diagnostic quality. DICOM ready system alongwith stand alone image compliant latest version PC with optimal processor speed and minimum 80 GB hard disc supported with complete networking. For this purpose the supplier would be required to supply a minimum six image compliant PC's equipped with necessary software e.g. reporting and analysis terminals to enable access and analysis of the patient images from different locations of the department

· Gantry: · ·

Vendors to provide a comprehensive list of users of dual head SPECT gamma camera in India and their performance profile

Gantry design should be wide open to avoid claustrophobic imaging with clockwise and anticlockwise movement. Height, width and depth should be adequate to conveniently locate the gantry in existing space available in the department. In order to access this, tenderers are required to mention the dimensions. Axis of rotation: 100-104 cm. Distance between SPECT & CT field of view: 130-136 cm. Weight with high energy collimator: not more than 3700 kg. Patient opening for high energy collimator (HE): Max 66 cm and Minimum: 12cm. Patient opening for low energy high resolution (LEHR) collimators: Max: 70 cm; Minimum: 19 cm. Patient positioning system: LCD colour display minimum of 15 inches flat panel. Auto contouring: The detectors should be equipped with automatic body countering (ABC) both for 90° & 180° SPECT with an average auto contour distance of 1.1 cm. Gantry should have emergency stop buttons. LCD persistence scope mounted on the gantry for continuous display of patient position and gantry parameters Max radial and lateral speed: 72 cm/ min Max lateral position: 11.9 (left) / 10.2 (right) Max CW/CCW: 365°/180° Ring position: 540° Rotation precision: 0.1° Rotation speed: 003-3.0 rpm

· · · · · · · · · · · · · · ·

· · Detectors:

Centre of rotation (COR): 0.25 pixel in 64x64 matrix Caudal tilt: Max: ±16°

Dimensions: · · · Large field of view with UFOV of 52 X 40 cm or more to enable adequate patient breadth coverage and diagonal field of view of 63 cm or more. Should have facility for automatic correction for energy, linearity and uniformity Integrated CT hardware option for transmission attenuation correction and lesion localization for all applications

Crystal: · · · Size: 59 X 44 cm or more Diagonal: 69 cm or more Thickness: 9.5 mm (3/8").

Photomultiplier Tubes: · · Number of PMTs: 55 or more per detector with 1 ADC per PMT (True digital detector) Hexagonal array with box type dynodes and high overall efficiency.

Shielding characteristics of Detector: · · · · Back: 9.5 mm Sides: 12.7 mm Patient Direction: Max ­ 36.4 mm; Min ­ 27.9 mm. Edge to edge: Detector housing and FOV ­ 7.6 mm.

Intrinsic Spatial Resolution: · · · · FWHM in CFOV: 4 mm FWHM in UFOV: 4 mm FWTM in CFOV: 8 mm FWTM in UFOV: 8 mm

Intrinsic Energy Resolution: · FWHM in UFOV: 10 %

Intrinsic Flood Field Uniformity: · · · · Differential (CFOV): 2.5 % Differential (UFOV): 2.7 % Integral (CFOV): 3.0 % Integral (UFOV): 4.0 %

System Spatial Resolution (LEHR at 10 cm): · · FWTM in CFOV: 7.8 mm FWTM in UFOV: 14 mm

Intrinsic Spatial Linearity: · · · · · · · Differential (CFOV): 0.2 mm Differential (UFOV): 0.2 mm Absolute (CFOV): 1 mm Absolute (UFOV): 1 mm Multiple window spatial registration: 1.0 mm Intrinsic maximum count rate: 310 kcps Intrinsic spatial resolution at 75 kcps FWTM (CFOV): 4.6 mm FWTM (UFOV): 9 mm · System Planar sensitivity (LEHR at 10 cm)

Absolute: 200 cpm / µCi or more · System planar sensitivity (MELP at 10 cm) 400 cpm / µCi or more · Reconstructed spatial resolution with scatter (LEHR): 12.00 mm

Collimator Specification: · The tenderers are required to provide following collimators along with collimator exchange cart for each set of collimators 1. Low Energy High Resolution (LEHR) 2. Low Energy Fan Beam (LEFB) 3. Medium Energy General Purpose (MEGP) 4. Pinhole (Single) 5. High Energy General Purpose (HEGP) - Optional to be quoted separately

CT sub system specifications: Tube Details: · · · · · · · · · Tube current: 20-345 mA Tube Voltage: 80, 110 & 130 kV Heat storage capacity: 6.0 MHU Anode heat storage capacity: 4.2 MHU Focal spot size: 0.8 X 0.7 mm Rotational time: 0.6, 0.8, 1.0, 1.5 second etc. Temporal resolution with heart view CT option: 150 m Second Maximum scan time: 100 seconds Power generator: 48 kW

Filter Assembly: · · Al-Equivalent: 6.3 mm Beam limiting device: 0.5 mm

Acquisition System: · · · · No. of rows of detector: 6 No. of channels per slice: 1344 or more No. of projections: 2500 (1/360°) Maximum no. of slice per rotation: 06

Gantry Aperture: · Patient part diameter/ aperture: 70 cm

· Scan field: 50 cm Data Sequencing: · Tenderers are desired to furnish complete details of the data sequencing including reconstructed slice widths, full scan times (360°), partial scan time (240°), no. of uninterrupted scans per range, auto range, standard scan cycle time, dynamic scan cycle time etc. In addition tenderers would also be required to provide details of multislice spiral image reconstruction slice specifically with reference to reconstructed slice widths, slice increment length and reconstructions time, topogram length and possible views.

Image Quality: Low contrast resolution: · · · Object size: 3 mm Contrast: 3 HV Surface dose: 20 mGy Air Kerma or less at 100 mAS; 16 mGy Air Kerma or less at 90 mAS using parameters such as 0.6 second, 10 mm and 130 KV

High Contrast Resolution: · 0% MTF: 17.5 lp/cm

· ·

2% MTF: 15.1 lp/cm at 60 mA, 130 kv, 0.8 seconds and 1.0 mm

Homogeneity: · · · Cross field uniformity in a water phantom: max± 4 HV Dose (CTDI100 volume): Phantom 16 cm 32 cm Image Display: · · · · · Range of slice thickness: 0.63 ­ 18 mm Scan field size: 50 cm Reconstruction field size: 5 ­ 50 cm Reconstruction matrix: 512 x 512 Houns field scale: -1024 to +3071 13.8 mGy/100 mAS 110 kv 130 kv

20.0 mGy/100 mAS 6.3 mGy/100 mAS

04.1 mGy/ 100 mAS

Patient Bed specifications: · · · · · · · · Length: 240.0 cm or more Width: 80 cm or more Height: 130 cm or more Range of vertical motion: 54 ­ 120 cm Pallet material: Carbon fiber Pallet thickness: 15 mm or less Pallet width (SPECT) 40.0 cm Attenuation at 140 keV: less than 10%

· · · · ·

Maximum weight bearing capacity: 205 kg Maximum deflection of patient pallet: less than 2 mm Scan length in Whole body mode: 200 cm Horizontal range: 33 cm with an accuracy of 0.5 mm Adjustable head holder for brain scan, butterfly arms support including cardiac arm support, leg support, cushion pad and wide straps for decreasing breast attenuation for optimized cardiac SPECT studies. Pediatric & scintimammography pallets. Separate pallet for imaging to enable radiotherapy planning. System compatible ECG Trigger to be provided with two sets of spare leads.

· · ·

Acquisition Unit: · One acquisition station independent of main processing unit capable of data acquisition in static, dynamic, multi-gated, list, whole body scanning, SPECT and gated SPECT with facility for anatomical markers Acquisition consol should allow universal networking via DICOM ready to both local and wide area networks Display of the alpha numerical patient acquisition data High performance Intel dual core PC with multi tasking operating system having minimum of 2 GB RAM, 3.0 GHz processor speed, 200 GB SCSI hard drive and high resolution flat panel LCD monitor of minimum of 17" size. It should also have CD and DVD combo drive with writer facility. Fully integrated CT system capable of acquiring X ­ray transmission data along with nuclear emission data. SPECT & CT data acquisition should be on the same console. The CT should have multislice capability with 4 or more slices in a single rotation Image acquisition and data display should be from 64x64 matrix up to 256x1024 matrix. Acquisition termination by preset time, preset count or manual stop Pre-defined acquisition protocols as well as facility for user to configure his own customized protocols Should provide ECG gating device with R wave display

· · ·

·

· · ·

· ·

Virtual film sheet with image manipulation. The acquisition workstation should be DICOM ready to permit-

(i). Exchange of images and other information (ii). Communication with other manufacturer's equipments/work stations. (iii).Workflow with hospital information system and other radiological information systems. (iv). Coverage for access controls of patient multiple data CD writers. (v). Maximum intensity projections, multi-planar reconstructions, various display format in routine use. (vi).Image fusion volume rendered CT data fused with SPECT data in a single image frame with 3 x 3 display. Processing & Software workstations: Four nos. (for various locations of the department . Out of four one work station should be along with LCD multimedia projector to be installed in teaching room): · · · · · · · High performance Intel dual core PC with multi-tasking workstation with full DICOM ready with image transfer print etc. Minimum of 2 GB RAM, 3.0 GHz processor speed and 200 GB SCSI hard drive logically divided in to 3-4 partitions Minimum 21 inch high resolution LCD monitor 3.5" FDD and standard DVD & CD combo drive for data archiving and retrieving with write facility for both and 1000 DVDs, two 2 GB flash (pen) drives also to be provided The computer is to be connected via a DICOM network for processing and storage of the data to the already existing processing and acquisition computers and documentation devices Broad band remote diagnostic facility to be provided The software, apart from other state of the art applications, should also provide following applications

(i). Real time display of volume data set and enable automatic quantification of stenosis and evaluation of aneurysms. (ii). Clinical processing software and comprehensive protocols for wide SPECT (CT). spectrum of SPECT,

(iii) SPECT reconstruction, automatic cardiac processing, motion correction, whole body and whole body SPECT. The software should also encompass organ specific protocols for kidneys, lungs heart, thyroid, parathyroid, brain gall bladder, liver, osteology, oncology etc. (iv). 2D & 3D volumetric visualization and quantification for assessment of myocardial viability and perfusion and correlation with angiography. (v). Quantitative evaluation and display of 2D/3D gated as well as nonperfusion. (vi). Standard package of CT software including those for angiograph (cardiac and renal) applications and other navigation applications. (vii).Provide QGS/QPS cardiac software with 2D/3D display and review of wall motion myocardial perfusion analysis. (viii).Should provide evaluation of CT images, coronary calcification display of dynamic CT data. · · · (calcium scoring), gated myocardial

Apart from inbuilt SPECT/CT software for fusion, separate software for fusion of imported CT and MRI data with SPECT is also to be provided Software for various scatter corrections and filters, standard attenuation correction with CT data , patient motion correction etc Various clinical application softwares including SISCOM, SEGAMI, NEUROGAM on NEURO MATCH and SPM2 etc.

QC Utilities: · · · · · · Ba-133 point source and Co-57 point source

Fillable flood phantom for rectangular field of the size adequate for the gamma camera to be supplied. Co-57 flood source of at least 15 mCi strength (on the day of the delivery to the institute) for rectangular field of the size adequate for the gamma camera to be supplied CT quality assurance phantom for contrast resolution, radiation safety, image uniformity and pixel noise etc. QC software (CT) for verifying alignment of the table position between SPECT and CT acquisition SPECT phantoms, point sources, CT phantom & KV/mA meter etc to permit comprehensive quality assurance program both for SPECT & CT

· · · · · ·

Four Quadrant BAR phantom for rectangular detector size compatible with the detector systems size of the camera to be supplied Intrinsic and System resolution phantom System count rate performance phantom Brain phantom and liver slice phantom having cold and/or hot lesions of various sizes QC software package (NEMA NU2-2001 or latest protocol) with documentation Anthropomorphic phantom (optional)

Accessories: · · · · · · · · · · · · High resolution table top dry laser film processor for 8" x 10" and 14"x 17" (CT size) x-ray films with minimum 100 packets each. photo quality laser colour printer Future up gradation of the software or new developments shall be required to be done by the vendor free of charge from time to time. Clinical programming language for user programming Video imager with networking (optional) Tread mill with 12 lead ECG display and processing and a hard copy out put device Capintec Thyroid uptake system with trolly and computer as well as data storage facility and out put device. Infusion pump with 5 to 50 ml range syringe capacity Two Capintec radioisotope calibrator for beta and gamma activity measurement (CRC-15, item no. 5130.3097). Multi-channel Well scintillation gamma counter with simultaneous 10 tubes counting facility with computer for energy range from 35 keV to 667 keV with UPS. 2 Rotem ion meters and 2 Rotem survey meters for monitoring gamma radiation. 3 Aloka pocket dosimeters (PDM-112)

· · ·

Separate Optical data storage device capable storing a large number of clinical studies data for a large no of studies. Syringe carrier and holders (six each). System to be supplied with online digital UPS of separate make of appropriate capacity providing 30 minutes backup with maintenance free batteries

Others: · Comprehensive warranty of five years. Warranty of the equipment shall include crystal, CT tube and all accessories (Thyroid uptake, Well counter, Isotope calibrator, Gun monitor, Survey monitor, Pocket dosimeter, Laser printers etc) as well as leads, wires, electronic consumables, third party items to be supplied by vendor. CAMC (Comprehensive Annual Maintenance Contract) for a minimum of five years to be quoted separately (CAMC would include all kinds of spares that might be required for insuring 95% uptime of the equipment CAMC shall also include the crystal, CT tube and accessories such as Thyroid uptake, Well counter, Isotope calibrator, Gun monitor, Survey monitor, Pocket dosimeter, Laser printers etc and other items as mentioned above. Cordoning off of the acquisition terminal with lead and appropriate sized lead glass. Minors modifications in the walls, floors and other areas of the room to permit adequate shielding particularly for CT associated radiation safety requirement. Thickness of the walls/ doors to insure safe radiation levels in the working areas around the Gamma Camera room. Minimum 3 training slots of four weeks duration at an advanced centre abroad (to be approved by the institute). On site application training in phases to technical and clinical staff initially and also on as and when required basis. A fully trained local resident engineer to be posted to ensure optimal uptime and effective and prompt maintenance services subsequent to satisfactory installation. The tenderer would have to submit their offer as buy back of the existing gamma camera on "as is where is basis" which can be inspected during any working day. The value of buy back offer should be quoted separately by the vendors. The acceptance of the installation shall be subject to satisfactory handing over of the system to the department and certificate to this effect to be issued by the institute. All the operating and service manuals in duplicates to be provided by the vendor at the time of handing over the machine.

·

· ·

· · · ·

· ·

Specifications for ELISA Reader with printer spectral range 400-750 mm, read out range 0-3.5 Abs., accuracy: ± 1% (0-0.2 abs) precision: cv <0.2% (0-2.0 abs), reading speed 5 seconds entire plate, supplied with minmum 4 interference filters (405,450,492,620 mm), Application: absorbance reading of 96 well microplate,

Specifications for ELISA Washer Wash/waste bottle capacity : 2 liter , Operating pressure 0.5 bar, Residual volume 0.5 µl, Precision 5%, Operating cycle continuous, Wash head 8 and 12 way, Programmable volume washes, soak time and pause time. Application: Washing microplate during ELISA procedures. Specification for Gel-Documentation System Gel-Documentation System (Gel doc) 12-bit CCD Camera with Anti-reflective coating and MicroLens Sensitivity upto pg level Full screen refresh rate of atleast 30 frames per second for real time viewing Motorized Zoom Lens Darkroom should be chemiluminescence imaging ready Should have Filter Wheel for Multiplex Imaging Dual wavelength UV Transilluminator (302/365 nm without changing tubes); Imaging area of 21x26 cm, Mounted on slide out tray for easy access Independent source for Tranilluminated White Light; imaging area of atleast 21x31 cm Dual Epi-White Lights Software for acquisition, enhancement, editing, annotation, archiving & analysis including features like I-D multilane densitometry, 2-D spot densitometry, MW, Rf analysis, Mass calculation, Microtiter plate, Eli-spot, Array & Dot Blot Analysis, Colony, Cell & GFP Yeast counting, Q-PCR Zymogram gel analysis, Band maching, RFLP, RAPD, Fingerprinting, Dendrogram creation, options for Dice, Jacard, Pearson, Frequency, Similarity Coefficients & Cluster Analysis with multiple methods including Neighbor joining, UPGMA, WPMA, Simple Linkage, Complete Linkage, Ward, Median, Centroid etc., Gel Scoring, Multi-color fluorescence microscopy imaging & Movie Mode facility. Computer (with Intel Pentium-4 processor @ 2.93 GHz, 80GB HDD, 512 MB RAM, CDRead/Write, 17" Color Monitor, Keyboard, Mouse)

Specification of Hot Air Oven

Hot Air Oven

· · · · · ·

High temperature Oven (ca-110-120 liters) with stainless steel interior, two shelves and high heat safety device with TWB class 2 standard. Uniform heating with dual mode by convection and digitally controlled air unit. Temperature fluctuation should not exceed 0.40C Temperature range: ambient +50C to 3000C with adjustable heat load (0 to 100%) Microprocessor controlled operations with LED display. The set and delivered parameters should recordable to meet the GLP standards Internationally accredited instrument with ISO 9001. The quotes should be substantiated with proper catalogue Three years warranty with AMC (comprehensive).

Specification for Millipore water purification system (Buy Back): (specifications enclosed) Sl.No. Name of Technical Specification equipment Approx. Expected Rema cost (Rs) Qualitative and rks, if (Lakhs) quantitative any benefits to the Department 9.00 Department is involved in the molecular and cytogentics diagnostic work for which the quality of water is very important component. For this purpose we need Milli Q system.

1

Milli Q system consisting of Elix & Milli Q (1)

Two stage purification 5 micron and 1 micron polypropylene graded desnity wrapped type depth filter with low voltage 20 watts powered DC pump with noise levels of 50 bp. Product water should have Resistivity of 1015 Megohm, TOC levels less than 30 ppb and Bacteria level of < 1 cfu/ml, Flow rate 8-12 litre/hour sgould have 3 stage

purification step. Stage 1: Pretreatment cartridge with antiscaling compound and silver impregnated. Stage 2: A high flux thin film RO membrane with 100 Dalton cutoff and compatible with feed water quality of SDI upto 20 and free chlorine level of 3 ppm and conductivity of 2000us. System should have a conductivity meter before and after RO membrane to monitor. Stage 3: Self regenerating Electrodeionisation module with mixed bed ion exchange resin shpuld have carbon bead at cathode and should not have conditioning cartridge before EDI. Monitoring : coaxial resistivity cell with cell constant 0.01 cm-1 & autotest resistivity measurement. Unique temperature feed back pump with temperature compensation to 0.1o C for consistent flow rate irrespective of temperature changes. System should have an in built display to ensure the system operating parameters are displayed a all

times. System should have Auto diagnostic (facility with Error No and Alarm code for easy Blow molded cylindrical conical bottom PE reserovlr with 30 L capacity with sensor rod float switch and single 3 stage vent filter consisting of soaline and activated carbon and 0.65 micron hydrophobic membrane and have the option of using automatic sanitization module toprevent the bacterial growth and biofilm formation. Final product water having Resistivity of 18.2 Meg ohm. Cm, TOC 5-10 ppb, Bacteria < 1 cfu/ml, Pyrogen < 0.001 Eu/ml is delivered by point of use dispenser with rocker arm, volumetric dispensing and auto shut off facility. Stage 1: sould have separate feed water specific purification cartridge and application specific polishing cartridge. The cartrdige flowpath should have UPFLOW in purifcation step and DOWNFLOW in polishing step. Cartridge must remove volatile

organicsa application specific cartridge both must remove volatile organics. Step 2; Should have a 5000D cutoff UF cartridge with UV between 5.6 and 7.65 over pyrogen challenge levels of 442 and 44200 EU/ml. Stage 3: Final filter 0.22 micron PVDF, stacked disc validated membrane. System must have a Footswitch for Hands free operation. Monitoring : System should have a provision to install built in TOC monitor with separate Uvlamp having optimum oxidation cycle for complete photo oxidation of organics and should mesure the TOC range in the unit of 1 to 999 ppb confirming to USP 26 suitability test requirement.

Specification for Roto Sealer Sl No 1. Name of item with fill specification Rotary Crimp Sealer with table & printer for closing sterilization pouches Qty. (in figure and words) Two Only Cost per unit (approx.) in RS. 02 Lac Total cost (Approx.) 04 Lac

Specification: Seal Speed : 9.8m/Min. Seam width : 15 mm Sealing temperature - 100 ­ 200º C with facility for pre selecting the temperature. Main Voltage 220V Net weight 19 ­ 20Kg. Sealing unit 540 x 325x19.5mm Table (SS) 740x192x100mm. Roller conveyor (SS) - 755x285x76mm. The printer should print in two sizes packaging date, Expiry Date, lot number and user code no. on paper side of pouch in one line. Low noise, maintenance free electronic temperature control, digital temperature indicator, digital counter 03 digits, Automatic start/stop of conveying system . Specification for Shaking Water Bath Microprocessor Controlled Shaking Water Bath · Temperature range: +100C to 99.90C, approx. 50C above ambient to 99.90C, approx. 30C above tap water temperature (1083: after installation of level regulator 1919, accessory) to +99.90C Temperature display & setting: digital-LED Temperature constancy (temporal): ±0.10C Shaking motion: reciprocating, optional Shaking amplitude: 22mm Over temp. cut-out: Electronical, 40C above set point and Electro mechanical> 1300C Shaking frequency: 10-150 min-1 Shaking frequency setting: stepless Electrical connection: 230 V /50/60 Hz /1.5 kW Three years warranty AMC (comprehensive). Interior dimensions: 450x300x210mm (WxDxH) Clamps for flask 25ml (50), 50ml (30), 100ml (20), 250ml (15), 500ml (10) Test tube racks: 160x16/17mm, 100x17mm (200 tubes), 100x12mm (350 tubes) With adjustable water level device. Specification of Syringe Infusion Pump 1. Capability of working on AC power and internal rechargeable battery lasting over 3 hours with full charge. Battery should be easily replaceable and should have provision for running on 12V DC supply like ambulances.

· · · · · · · · · · · · ·

2. 3. 4. 5. 6. 7. 8. 9. 10.

Should be capable of using 10ml, 20ml and 50ml syringes of more than 5 commonly available commercial syringes. Delivery rate should be from 0.1 ml/hr to over 100 ml/hr. Selectable purge rates for different syringes. High occlusion pressures in multiple settings should be able to deliver in arterial lines. Pump operation should be microprocessor controlled with accuracy of + 5%. Alarms for occlusion, near empty, low battery etc., at different settings. User-friendly operation knob. Optional capability for setting the proper flow rate when dose rate, body weight etc are given, and capable of changing from ug/kg/min to mg/kg/h. Light weight, portable and capability for pole mounting. Separately quote for angle adjustable pole clamp.

Specification for Gradient Thermal Cycler Gradient Thermal Cycler Universal block 0.2 ml (96 Capacity), 0.5 ml (77 capacity), Gradient block temperature range 1-20°C, Peltier technology based, Temperature control range 4-99°C, block homogeneity 20°C to 72°C: ± 0.4°C, Temperature control speed 2 to 3°C, Heat able lid, Number of cycles 99 Regulation Accuracy per well +/- 0.20C Temperature control speed approx. 30C/s (heating) Temperature control speed approx. 20C (cooling) Number of programs 100 in device Auto calibration in tandem with temperature validation system Log book function for error messages and new calibration

Specifications for Transilluminator: UV Transilluminator with filter 312 nm, Size 20x20 cm Specification for Automatic Tissue Embedding Centre Name of item with full specifications Automatic Tissue Embedding Centre (On buy back basis) Specifications: Complete paraffin embedding system. Preferably 2-3 pieces modular unit including

paraffin dispensing unit with hot plate, fast achievable temperature of cold plate and mould/cassette storage unit with following features: Digitally controlled touch screen display. Paraffin tank 3-5 liter. Cold plate to hold 60-80 standard size moulds. Peltier controlled cold spot for specimen orientation and making blocks for stat work. Easy drainage and disposal of melted wax. (Drained wax should remain in melted form) Hand & pedal switch for control of paraffin flow rate Magnifying glass with cold light source. Preferably in-built heated forceps module for different size of forceps tips. Large tissue storage area. Cold plate temperature - -5 to -15 degree Celsius Hot plate 55 to 75 degree Celsius Tissue storage 40 to 80 degree Celsius Mould storage 50 to 75 degree Celsius Paraffin tank 40 to 75 degree Celsius Servo voltage stabilizer of appropriate rating Moulds of different size ­ 80 number Three years comprehensive warranty Power requirement : 220 ­ 230 V, 50/60 MHz Mention items included with equipment Also mention optional accessories separately SPECIFICATION OF (DIGITAL) INTRA VASCULAR ULTRASOUND SYSTEM (IVUS)

-

Automated border contours- fast, simple image interpretation Automated lumen and vessel measurements to aid in diagnosis and planning Colorized tissue map of plaque composition ­ for complete lesion assessment Dual 3.06 GHz Xeon workstation ­ class computer At least 80 GB EIDE Hard drive 2 GB SDRAM or more 16x DVD Drive ECG trigger card Digital frame grabber Full color printer

Approval International standards setting agency : US FDA or equivalent

Monitor S-VGA monitor (17" or more) Data entry, Key board, track ball and point & click interface with coded colour keys Multiple image screen formats : Ultrasound, fluoro, stored and sagittal views Multiple image diameters : Coronary 8mm to 16mm, Peripheral (small) 8mm to 24mm, Peripheral 20mm to 60mm Patient data entry and on screen annotation Automatic and manual measurements : Four diameters and two areas with % differences Biophysical inputs ECG Digital videoloop storage at least upto 8 minutes with eight still frames DICOM storage to CD-R and PACS network VCR storage ChromaFlo and In-Line Digital Display Vendors should quote with compatible IVUS Catheters.

Specifications for ICU Ventilator Basic Description: Positive Pressure, Invasive, Conventional Patient Population Range: Neonate, paediatric, and adult Purpose: All purpose, Bed side + Transport Ventilator Class: Intensive Care Unit (ICU) Power: AC 220V/60Hz and Battery back up Dockable oxygen-gas trolley Modes: Pressure control (PC) Volume control (VC) Pressure Regulated Volume Control (PRVC) Syncronised intermittent mandatory ventilation (SIMV) Pressure support (PS) SIMV + PS CPAP Automode Open Lung Modes/Tools Waveforms:

Rectangular, Descending, Sinusoidal Operating Range of Controls: Respiratory Rate: 0.5 to 70 bpm Tidal Volume: 0.10 to 2.5 liters Peak Flow: 10 to 120 Lpm O2%: 21% to 100% Trigger Sensitivity: 0.5 to 20.0 cmH2O Inspiratory Pressure: 5 to 60 cmH2O above PEEP I:E ratio: Inverse ratio up to 4:1 Inspiratory Time: 0.2 to 9.9 seconds PEEP/CPAP: 0 to 40 cmH2O Pressure Support: 1 to 70 cm H2O Graphic Display: Tidal Volume,Set Rate (bpm),Actual Rate,Peak Flow Lpm,O2% CO2 waveform,PEEP/CPAP,Resistance ,Compliance,PV loops Flow-volume loops,Other curves LCD screen clearly displays breaths as a waveform or loop for in-depth analysis Humidifier: Draw-over, heated wire. Non technical requirements: 1. Quote the price of all optional modules separately 2. On-site Warranty: Three years Yearly cost of comprehensive annual maintenance contract for post warranty five years. The cost of AMC will be used for cost comparison purpose. Remarks ICU expansion is impossible in the absence of new ventilators

Specifications for Chemiluminescence Reader

1. Fully automated Chemiluminescence Immunoassay system ­ One No. · · · · · · · · Random Access Analyzer On Board Washing Facility On Board dilution facility Auto clot detector Each available analyte will be assessed on the basis of (submit in a table form) Sample Type (serum/plasma/urine/CSF), reporting time, standard range, sensitivity & pack size availability. Product insert for each analyte be provided. List of Rapid/stat Assays and Third Generation Assay be provided. Should be able to provide maximum numbers of analytes as mentioned below

ACTH, Cortisol, Ferritin, Folic Acid, Vitamin B12, Red Cell Folate, Calcitonin, Osteocalcin, Intact PTH, 25(OH) Vitamin D, 1,25(OH)2 Vitamin D, Urinary Albumin, CPeptide, Insulin, Growth Hormone, IGF-I, IGFBP-3, High Sensitivity CRP, Androstenedione, DHEA-Sulfate, Estradiol, FSH, LH, Progesterone, Prolactin, SHBG, Total Testosterone, Free T3, Total T3, Free T4, Total T4, TBG, Thyroglobulin, Anti-TG Ab, Anti-TPO Ab, Thyroid Uptake, Rapid TSH, Third Generation TSH, PSA, Cyclosporine, Aldosterone, Plasma Renin Activity, Homocysteine, Bone Markers. · · · · · · · List of other available assays be provided. Electrical requirement (UPS be provided along with the equipment). Any other requirement. List of parts which are needed to be replaced on regular basis. Warranty -Two years. Rates of the kits should be freeze for five years. AMC (comprehensive) should be signed at the time of order.

Specification for Deep Freezer Name of item with full specifications Deep Freezer (Chest type) Temperature range ­ 80 C Dedicated for plasma storage in blood bank 450 ­ 500 lts capacity Microprocessor control for operation LCD display of temp. variation Co 2 backup, temp chart recorder, preferably water-cooled condenser Pt sensor for connection to external monitoring system preferred Deep Freezer (Upright) Temperature range - 20 C Double door Dedicated for plasma storage in blood bank 500 ­ 600 lts capacity Microprocessor control for operation LCD display of temperature, CFC Free Audiovisual alarm for temp. variation CO 2 backup, temp chart recorder Pt 100 sensor for connection to external monitoring system preferred Specifications for Deep freezer -40o C

Upright Model -40 Degree C, Gross Capacity 310 Ltr, Temperature Range -30 degree C to -40 degree C, Electronic Display +/- 1 C digits, High Temperature Alarm, Door Alarm Noise level 48 dba, All round sealed drawers to prevent frost escape when opening door Power supply 230V, Equipment supplied with 7 drawers, Voltage free contact for remote monitoring , Quick freezer button Display lock on/off function Specifications for Deep freezer -20o C · Temperature control range ­20 to 300C · Capacity 274 liter · Basket 4 large & 1 small · Cooling system hermetic · High low temp alarm · Power failure alarm · Microprocessor temp control · Memory backup · Digital temp display · CFC free refrigerant Vertical model

Specifications for Intra-operative Ultrasound with colour Doppler Name of item with full specifications Intra-operative Ultrasound with colour Doppler (against buy back machine make ATL model LMHA Sl. No. X 11342 with switchable probe ­ 5, 7.5 and 10 MHz) Specification: 1. High resolution digital 3D ultrasound with 17" LCD display monitor. 2. Probe 7.5, 10 MHz both linear sector & Laparoscopic 3. Color Doppler 4. Digital Video recorder on DVD 5. Coloured Printer 6. Necessary spares, cables & accessories. 7. DICOM (Version 3.0) ready

Specifications for Miscellaneous Laboratory Equipments Microscope, Microscope: The instrument should be sturdy, fitted with plan achromatic

objectives plan 4x, 10x, 60x (spring loaded) and 100x (spring loaded) on a reversed quintiple nosepeice with click stops. The optical system should be color corrected for infinity with antifungus property built in transmitted Koehler illumination. The microscope stand should have co-axial focusing knobs for coarse and fine adjustment with upper limit stopper. Wide field high point eye piece 10x, 22 mm with diopter adjustment (+2 to -8) and rubber eye shield (pair) with interpupillary distance of 48 to 75 mm. Binocular head inclined at 30-45 degree with 360 degree rotation. Rectangular scratch resistant stage with right hand control with double slide holder and vernier calipers on X Y axis. Plan achromatic universal type condenser (dry type) with numerical aperture 0.9-1.2. Transmitted light filters for day light and neutral light. Illumination - Minimum 6 V, 20 W quartz halogen lamp. Power - 220 + 10 V, 50 Hz. Vinyl dust cover. All the necessary adaptor and power cord should be provided for functioning of microscope. Five additional halogen lamps should be provided. Instruction and operational manual. Electronic Balance Electronic Balance: High precision digital electronic balance from reputed manufacturers. Capacity 1500 gm. Readability 0.1 mg. Internal calibration. Weighing mode in grams/milligrams. Easy access door. 200 V, 60 Hz. Soft touch pad. · · · · · High temperature Oven (ca-110-120 liters) with stainless steel interior, two shelves and high heat safety device with TWB class 2 standard. Uniform heating with dual mode by convection and digitally controlled air unit. Temperature fluctuation should not exceed 0.40C Temperature range: ambient +50C to 3000C with adjustable heat load (0 to 100%) Microprocessor controlled operations with LED display. The set and delivered parameters should recordable to meet the GLP standards Internationally accredited instrument with ISO 9001. The quotes should be substantiated with proper catalogue

Hot Air Oven

Three years warranty with AMC (comprehensive). Incubator Microcentrifuge Microcentrifuge: Refrigerated fast cool -9 degree C ambient, CFC free large LED display and soft touch keypad. Speed upto 14000 rpm and 20000 g fast acceleration to maximum in less than 15 second. Brushless whisper quiet motor. Timer setting upto 99 minute, Hold function, quick release rotors with 24 1.5/2.0 ml microtubes. Rotor for 0

Centrifuge (imported & Indeginous) Table Top Centrifuge Table top centrifuge: Maximum speed 10000 rpm/7, 176xg Complete with 12x1.5/2.0 ml rotor and 0.5 ml adapters Ideal for cell separations and microfiltration Extremely quiet and cool running Safe for cold room use. · · · · · · · · · · CO2 Incubator Microprocessor controlled refrigerated micro centrifuge with 3 programmable memories Maximum RPM 18,000 and maximum RCF 31,514 with input in increments of 10 Rotor imbalance detection through out run time with imbalance switch-off Lid locking, holding and lid dropping protection must be standard with emergency lid lock release Brushless maintenance free drive for extra ordinary speed stability Aerosol tight rotors and automatic rotor recognition system Impulse key function for short centrifugations Dimensions (HxWxD): 280x330x620 mm) It should be ISO 9001 certified 3 years warranty with AMC comprehensive

Non Refrigerated Micro Centrifuge

APT line preheating chamber technology, Microprocessor controller MB1 for temperature and CO2 concentration, Communication interface RS 422 for standard software APT-COM data control system, Weld less deep-drawn, polished inner chamber made of stainless steel 1.4301/V2A or copper, Round edges, no fix installations ­ very easy to clean, Digital, drift free CO2-infrared measurement system with FPI difference measurement system, Fan-assisted air ­ jacket system, Hot air auto-

sterilization at 184.5 deg C for 1 hour, Gas mixing head ­ no fan in the inner chamber, Perm dry ­ system, condensation free double ­ basin humidification system, Electronic malfunction auto diagnosis system with visual and acoustic alarm and a potential free contact for central monitoring, Safety device (TWW) class 3.1 (DIN 12880) Acc. To DIN 12880, Inner glass door, tightly closing, 3 perforated shelves, stainless steel 1.4301/V2A house set for gas supply with hose clamps 230V-50Hz BOD Incubator Double walled, with the temperature range of +5 deg to 50 deg +- 1 deg, Thermostat controlled inside chamber and shelves of stainless steel/ Aluminium and outside of mild steel sheet, fitted with two air circulating fans and acrylic transparent door inside Size : 12 cu ft 900 x 650 x 580mm, Capacity: 336 ltr. Digital temperature controller cum temp indicator with Sensitivity of +- 0.5 deg · Microprocessor controlled refrigerated micro centrifuge with programmable memories. · Maximum RPM 18,000 and maximum RCF 31,514 with input in increments of 10 · Temperature range: -200C to +400C · Rotor imbalance detection through out run time with imbalance switch off · Rotors: (a) 24x0.2 ­2.0 ml rotor fixed angle RPM 18,000 RCF 31514 x g , (b) 6x 50 ml rotor fixed angle RPM 6,000 RCF 3,904 x g, (c) 24 x 0.2 ­2 ml swing out rotor RPM 13,000, RCF 18,000 xg · · · · · · · Lid locking, holding and lid dropping protection must be standard with emergency lid lock release Brushless maintenance free drive for extra ordinary speed stability Aerosol tight rotors and automatic rotor recognition system Impulse key function for short centrifugation Dimensions (HxWxD): 280x330x620 mm approx. It should be ISO 9001 certified Three year warranty with maintenance on the complete equipment with routine check-ups

Refrigerated Micro Centrifuge

Specification for NIBP equipment with multi-channel monitor Name of item with full specifications NIBP equipment with multichannel monitor against by back NIBP (Make MDE

ESCORT Model E 100 Sr. 10136 ) Specification: Bedside monitor having the following vital signs measurements options 1. Non-Invasive Blood pressure 2. CVP 3. Direct arterial line pressure 4. Oxygen saturation Heart rate, ECG, Pulse rate, venous flow. 5. CO2 concentration

SPECIFICATION FOR PLATELET AGGREGOMETER PLATELET AGGREGOMETER [quantity 1] Specifications Automatic optical aggregation system minimum two channel for aggregation tests in PRP or washed platelets with the following technical specifications: Should have sensitive optical detection device to determine difference between PRP and PP channels. Should involve low sample volumes of the order of 250 to 500uls. Should have facility for handling of Von Willebrand co-factor assays. Should have the facility to display the instructions for easy operation. Should be possible to run tests in replicate if needed with ability to monitor coefficient of variation between replicates. Should have ability to store standard curves for future use. Should be able to calculate test results in percentage of activity. Should be able to display real time growth including amplitude and slope data and possibilities to store and recall the test curves and the associated data. The system should be able to run with windows type of software for ease of operation and convenient data reduction. The machine should be quoted complete with software and the data reduction device including the necessary liquid handling systems and reagents.

General conditions for platelet aggregometer : 1. System should be quoted complete with appropriate stabilizer or UPS with one hour battery backup. 2. The recommended maintenance kit, for trouble free operation and maintenance of the system for a minimum period of three years should be quoted. 3. The consumables including reagents should be available for supply in local currency and preferably within 48 hours after ordering. 4. The after sales service should be available within 24 hours and with trained service personnel from New Delhi or nearer locations. 5. Reasonable installation base should be available in New Delhi/northern parts of India. 6. Installation and commissioning shall be done free of cost.

Specification Platelet Incubator with Agitator

Platelet Incubator with Agitator Capacity- 80 to 100 random donor platelet Storage Type of agitator: Flat bed agitation Continuous agitation with 60 to 70 agitation / min Shelves: Should be made of good quality Coated with bacteria resistant material Perforated so that air circulation on both side of bags Should be made of "non-slip" material Motor with internal fan Outside: Powder coated body with transparent Door with key lock Temp: Operating range 20 to 35 C 7 day chart recorder Temperature controller with temp. probe Refrigeration: Non-CFC air cooled refrigeration Safety features : Audio alarm for temperature fluctuation Auto stop for agitation when the door is Opened Lockable doors

Specification for Trans-Echocardiography System for Department of Anaesthesiology Cardiac Anaesthesia Services (Specifications for Echocardiography System:)

· Latest Technology all ­ Digital System suitable for ultrasound guided regional nerve blocks, Vascular Access, Adult Echocardiography, TEE Echocardiography, Vascular, Abdominal, TCD application. · Should have High Resolution non-interlaced monitor preferably TFT LCD screen. · System should have multi channel frequency compounding facility to filter low and high frequencies returned from the transducer and then process these signals independently & m parallel to create an image. · System should have 2D, M-mode, Colour flow, PW, CW, Steerable CW, and Directional Color Power Angio facility. · System should have minimum 128 channels · The system should support broadband Phased, Convex, Linear array TEE transducer technologies. · System should be light weight easy to carry for use in OT, ICU,CCU. · System should have minimum 256 gray shades. · The system should have an easy to use control panel, should have an alphanumeric keyboard with illuminated keys and status display which should be sanitizable while using in OT. · Cine review facility. · System should have preset for different type of applications special including nerve blocks. · Tissue Harmonic Imaging. · Should have compound imaging for better spatial resolution. · Intra operative, Multiplane TEE Transducer capability. · On-board Image management. · Storage-should have > 20000 image storage facility in the hard disk drive. · System should have inbuilt Image Management, with facility for direct storage of Images and cine-loop of 5-6 sec. · Archive-should have facility to transfer images to PC. · Should DICOM ready system with print, save, modality work list for connecting to DICOM network. · Full functional measurement facility and calculation should be possible.

With the following: · Broadband Phased Array Transducer for adult cardiac imaging. · Broadband Phased Array Transducer for Pediatric & Neonatal Cardiac imaging. · High Frequency, more than 10 MHz Linear Broad Bank Probe for regional anaesthesia nerve blocks, vascular access. Higher frequency will be preferred. · Colour laser Printer for printing of Images. · UPS of appropriate rating should be provided with the system. · Broad band sector Multiplane TEE Transducer. · Linear Array Transducer for Vascular imaging. · Intraoperative probe for Intraoperative applications. · All probes must be covered under warranty and during AMC. · All probes must be quoted separately.

TECHNICAL SPECIFICATIONS FOR VIDEOTELEMETRY SYSTEM The Video EEG System should have following specifications: 1. 40 Channel amplifier including 32 EEG Channels and 8 Polygraphy channels.The amplifier must have Ethernet technology. The channels should be sampled at more than 5,00,000 Hz simultaneously and to be stored at programmable rates of up to 1024 Hz. 2. CMRR of amplifiers should be greater than 110 dB. 3. Should have impedance check facility on the head box and on Recording PC. 4. Photic stimulator with LED's. 5. Should have possibility of review machine parameters. 6. Should have provision for notch (line) filter. 7. ADC resolution must have 20 bits or more simultaneous sampling. 8. The system shall have Pentium D Processor 2.8 GHz PC Dual Core with 2048 MB RAM, network card, multimedia speakers, mouse minimum 250GB hard disk for storing EEG for more than 120 hours continuously and shall operate under Windows XP and MS Office XP with Ethernet Network Interface Card. 9. Should be supplied with a high resolution 17" TFT Monitor and a laser printer. 10. Should be supplied with one camera with Zoom, PAN, Tilt for recording Video of MPEG in real time synchronization with Digital EEG. 11. EEG/Video Editing and Pruning of the EEG data. 12. DVD Writer for recording the Video EEG Data 13. Facility for exporting data using the software for review on any PC with any additional software.

14. The equipment must incorporate features like full optical isolation, anti-cautery devices and must confirm to international standards with certifications authorities. 15. There should be provision of two event markers 16. Report generation to be customizable and in MS Word format. 17. The Test information database software which should include information of patient database, resource scheduler and it should be customizable according to the end user. 18. Should have inbuilt pulse oximeter. 19. The electrode kit should consist of Gold Cup Electrode-100 Nos., EEG Conductive Paste ­ 10 Nos. & Skin prepping gel ­ 10 Nos. 20. Should be supplied with Ambulatory Recorder Capable of uninterrupted recording of minimum of 16 Channel EEG for two days with rechargeable battery pack and flash memory card. 21. The Software of Ambulatory recorder should be quoted as standard.

Specification of Non Contact Pneumatic Tonometer Specifications Console only dimensions Hand unit dimensions Console only weight Console with battery weight Hand unit weight Displayed scale Displayed accuracy Calibrated range Length of umbilical cord Electrical standards 355x305x205mm 14x12x8 265x115x40mm 10.5x4.5x1.5 3.18kg/7.0lbs 3.18kg/7.0lbs 0.887kg/1.95lbs mmHg +/-1 mmHg 7 to 50 mmHg 2.0m/6'6" I.E.C. 60601-1

Tender no. PGI/MM/ OPTH/33/06-07

SURGICAL OPERATING MICROSCOPE FOR NEUROSURGERY

Tender no. PGI/MM/ OPTH/34/06-07

1. Compact Microscope body with 1:6 zoom ratio & magnification range from 2.0 X to 18 X Continuously variable with 10x / 12.5x Eye pieces. 2. Dynamic ZoomLink for fast focussing at a low magnification & slow focusing at maximum magnification. 3. All the Optical Components should be apochromatic corrected for all three basic colours. 4. 5. 6. 12.5x wide field push in eye pieces Field of view should be 112mm to 11mm continuously variable. Spot illumination from 65mm ­ 11mm continuously variable

7. Variable objective lens from 200mm to 405mm or more continuously variable without changing the objective lens. 8. 0-180 Deg. Inclienable binocular tube with f= 200mm focal length. 9. Pair of handles( Removable) with easy control of zoom, focus & Four programmable buttons for Illumination ,Video control & on screen display etc.

10. Xenon illumination (180W Superlux) with Xenon backup for day light characteristics in the sturdy floor stand Should have easy touch button motorized balancing for Microscope body.

11. Mobile floor stand with on screen display showing total magnification, Working distance & other parameters like illumination, Video setting etc. 12. Stand should be capable of adjustment of focusing & zoom speed. Arm length should be about 1.75 mt. providing enough utilization space.

13. Integrated 1CCD Camera attachment with Speed/ Auto focus facility. The power supply of camera should be integrated in the microscope stand itself. With the facility of on screen display for Microscope parameters, Freeze/ live image, Rotate image & Mirror image. . 14. All the cable should be integrated in the stand for protection

Description of Horizontal High Pressure

HIGH VACUUM STEAM STERILIZER

ONE EACH. · QUANTITY: 24 CUBIC FEET AND 36 CUBIC FEET · SIZE: HORIZONTAL RECTANGULAR. · SHAPE: 600X900X1200 MM AND 900MMX900MMX1200MM · CHAMBER DIMENSION: ISI / COMPATIBLE TO INTERNATIONAL STANDARDS · STANDARD: 640 LTS AND 1000 LTS · CAPACITY: 121°C AND 134°C. · WORKING TEMPERATURE: 1.2KG. /CM2 TO 2.2 KG./CM2 · WORKING PRESSURE: DOUBLE OF WORKING PRESSURE · HYDRAULIC TEST: SINGLE RADIAL ARM · DOOR: 02 NO. AND 02 NO · CARRIAGE: THE STERILIZERS SHOULD BE CONFIRMING TO IS: 3829 WITH ISI MARK (VALID CERTIFICATE FROM ISI). SUITABLE FOR STERILIZATION OF WRAPPED AND UNWRAPPED S.S. INSTRUMENTS & OTHER DEVICES, LINEN, GLASSWARE, RUBBER GOODS &

LIQUID IN OPEN BOTTLES AND CAN BE OPERATED ON BOTH STEAM FROM CENTRAL STEAM FACILITY AND INBUILT STEAM GENERATOR.

Tender no. PGI/MM/ CSSD/04/06-07

(A)

(i) (ii) (iii) (iv) (v) (vi) (vii) (viii) (ix) (x) (xi)

MATERIAL OF CONSTRUCTION

CHAMBER: STAINLESS STEEL OF 316 OF IS 6911 OF 1992 JACKET: STAIMLESS STEEL. DOOR GASKET: HIGH TEMPERATURE RESISTANT SILICONE ELASTOMER TYPE SUITED FOR AUTOCLAVING OPERATION. MATERIAL SHOULD BE NON-TOXIC AND COMPLIES WITH THE REGULATIONS FOR AUTOCLAVING USE. INSULATION: NON FIBER SHREDDING RESIN BONDED GLASS WOOL WITH S.S. COVERING OF 304 OF IS 6911 OF 1992. DOOR HINGED TYPE WITH RADIAL ARMS WITH PROCESS LOCK TO PREVENT OPENING OF DOOR DURING THE PROCESS AND DOOR OBSTRUCTION SAFETY TO PREVENT CLOSING OF DOOR IN CASE OF AN OBSTRUCTION. MOUNTING: ON STAINLESS STEEL TUBULAR STAND WITH STURDY LEVELING LUGS. VACUUM SYSTEM: WATER RING TYPE VACUUM PUMP. THE SHELL AND TUBE TYPE CONDENSER SHOULD BE FABRICATED FROM STAINLESS STEEL OF 304. STEAM SUPPLY: THE STEAM GENERATOR SHOULD HAVE HEATERS OF REQUIRED ELECTRICAL LOAD, AUTOMATIC PRESSURE CONTROLLER WITH GAUGE, LOW WATER LEVEL CUT OFF, SAFETY VALVES AND WITH UTMOST SAFETY FEATURES .THE BOILER SHOULD BE OF STAIN LESS STEEL 316 WITH ARGON WELDING. STEAM TRAP: BOTH CHAMBER AND JACKET SHOULD HAVE STANDARD MAKE STEAM TRAPS. LOADING CAR (CARRIAGE): STAINLESS STEEL 304 GRADE OF IS 6911 OF 1992 WITH TWO SHELVES ENSURING A FLEXIBLE LOADING ARRANGEMENT. TROLLEY: THE TROLLEY FOR LOADING AND UNLOADING OF MATERIAL (ONE EACH AS PER THE STERILIZER SIZE).

(B) PROCESS

INDIVIDUAL STAINLESS STEEL CONTROL VALVES FOR PROCESS OF AUTOCLAVING AND ALSO FOR MANUAL OPERATION AS. (2) (3) (4) (5) (1) STEAM TO JACKET JACKET TO CHAMBER EXHAUST ­ FAST & SLOW VACUUM AND DRY DIRECT STEAM DISCHARGE LINE.

(C)

OPERATION OPTION (AUTOMATIC)

THE SYSTEM CYCLE SHOULD HAVE PLC BASED FRONT DISPLAY INDICATING THE FOLLOWING PARAMETERS. 1. (A) CYCLE NUMBER (B) LOADING INFORMATION (C) DATE (D) STERILIZATION DURATION (E) TOTAL PROCESS TIME (F) ALARMS FOR MALFUNCTION 2. CHAMBER AIR LEAK TEST. 3. PULSING AIR REMOVAL SYSTEM WITH DETECTION BY BOWIE DICK TEST. 4. REAL TIME DATA ­ REAL TIME TEMPERATURE AND GRAPH

OF TEMPERATURE INDICATION DEGREE CENTIGRADE VERSUS TIME IN MINUTES WITH SET POINT SHOWING THE SELECTED STERILIZATION TEMPERATURE ON STRIP CHART RECORDER. 5. RETRIEVAL OF DATA ­STORAGE OF ALL-IMPORTANT INFORMATION IN THE COMPUTER OF RECORD AND RETRIEVAL PURPOSE OF AT LEAST LAST 1000 ENTRIES. 6. PRINTER PROVIDED FOR HARD COPY.

(D) OTHERS

1. VACUUM PUMP:

MAKE HIGH CAPACITY WATER RING VACUUM PUMP OF STANDARD COMPANY OF REQUIRED CAPACITY COMPLETE WITH CONDENSER AND FITTINGS FOR MECHANICAL AIR REMOVALS I.E. VACUUM (PRE-VACUUM) AND DRYING (POST VACUUM). ANTI VIBRATION MOUNTING STAINLESS STEEL OF GRADE X 4 CR 19 NI 9 (ISS 304 SI) OF IS 6911 S.S. STEAM GENERATOR WITH AUTOMATIC PRESSURE REGULATION & OVER PRESSURE SAFETY & MEETING ALL NECESSARY REQUIREMENTS, AUTOMATIC FEED WATER SYSTEM. : ONE REMOVABLE PLUG SCREEN FOR CHAMBER DRAIN LINE TO BE PROVIDED : S.S. BAFFLE PLATE FOR EVEN DISTRIBUTION OF STEAM TO BE PROVIDED.

2. 3. 4.

NOISE CONTROL: ALL PIPES, FITTING, CONNECTION & STAND STEAM GENERATOR:

5. 6.

PLUG SCREEN BAFFLE

(E) TERMS AND CONDITIONS

ON TURNKEY BASIS I.E. ONLY WATER, ELECTRICAL SUPPLY AT SITE WILL BE PROVIDED. ALL CONTROL LIKE ICPT SWITCH, INPUT VALVES AND PRV IF REQUIRED SHOULD BE PROVIDED/INSTALLED BY THE SUPPLIER. 2. ISI CERTIFICATE AND LIST OF USER'S INSTALLATIONS OF RECTANGULAR STERILIZER MAY BE PROVIDED. 3. THE SUPPLIER WILL PROVIDE 2 SETS OF OPERATING MANUAL AND SERVICE MANUAL. 4. GUARANTEE/WARRANTY OF 05 YEARS. 5. OFFER FOR BOTH COMPREHENSIVE (LABOR & SPARE) AND NON-COMPREHENSIVE (ONLY LABOR) ANNUAL MAINTENANCE CONTRACT FOR FIVE YEARS AND PRICE LIST OF SPARE PARTS WITH VALIDITY OF RATES TO BE ENCLOSED WITH THE OFFER. 6. 95% UPTIME DURING WARRANTY AS WELL AS DURING MAINTENANCE CONTRACT. FINE/PENALTY AS PER MUTUALLY AGREED TERMS. 7. A COPY OF CERTIFICATE FROM BIS SHOULD BE ATTACHED. ALL TESTING CERTIFICATES FOR THE JACKET CHAMBER AND ALL OTHER COMPONENTS USED SHOULD BE PROVIDED. 8. THE TENDER WILL HAVE TO SUBMIT THE LIST OF COMPONENTS AND THE NAME OF MANUFACTURER OF COMPONENTS.

1.

INSTALLATION

9.

(F)

10. 11. 12. 13. 14. Safety 1. 2. 3.

FIRM WILL HAVE TO PROVIDE THE SERVICE REQUIREMENT DETAILS LIKE DEMAND, PRESSURE, QUALITY OF WATER & STEAM, ELECTRICAL LOAD, LINE SIZE FOR DRAIN ETC. THE FIRM WILL PROVIDE ISI NO. & APPROVAL FROM BIS. COPY OF CIRCUIT DIAGRAM, PLUMBING DIAGRAM, WIRING DIAGRAM, CONNECTION DIAGRAM ETC. PROVIDE TESTING CERTIFICATE/CALIBRATION CERTIFICATE OF FITTINGS, GAUGES, PID, PRV, VALVES ETC. HYDRAUTIC TESTING REPORT OF JACKET CHAMBER AND BOILER. CERTIFICATE OF MATERIAL OF CONSTRUCTION OF JACKET, CHAMBER, BOILER & DOOR.

THE

IBR APPROVED VALVES WITH GAUGES; THE TENDERER SHOULD PROVIDE TRAPS IN LINES AND SAFETY VALVES FOR JACKET AND CHAMBER FOR OVER PRESSURE SAFETY. DOOR SAFETY TO PREVENT STARTING OF PROCESS UNLESS THE DOOR IS CLOSED AND OPENING OF DOOR WHEN THE CHAMBER IS PRESSURIZED. INSULATED SURFACE TO AVOID SCALDING TO OPERATOR.

A High Speed Flow Cytometer capable of Cell Analysis and Cell Sorting is required which meets the following specifications

Tender no. PGI/MM/ HAEM/23/06-07

· A bench top, flow cytometer that can do cell analysis at high speeds and has a built-in high-speed cell sorter. · It should have the capability to acquire and analyze at least 70,000 cells per second. · It should be able to sort cells at least a rate of up to 25,000-30,000 cells per second with minimum purity of 98% irrespective of the number of lasers or colors used. · LASERS: Air-cooled solid-state lasers having wavelengths close to 488nm, 633nm and a 351 nm UV Laser of upto 20mW or above. There should be a provision for an optional 407nm laser. · It should have factory fixed and aligned optics for all excitation and emission paths; collection and excitation optics should not require daily/frequent alignment by the operator(s). · It should have the minimum capability of detecting 7 color florescence (i.e. 9 parameter instrument) which should be up gradable to 13 florescence (15 parameters). · The system should have advanced digital compensation with online as well as post-acquisition auto-compensation.

· It should be able to acquire samples from different size tubes (microtubes, 12x75mm, 15ml and higher capacity tubes). The system should have adjustable sample input agitation and temperature control system. · Must have 2- way, and proven 4-way sorting capability and the ability to assign sort modes independently to each sort stream. · The system should be able to sort cells in tubes, single cells into multiwell plates or microscopic slides. · Should be equipped with different nozzles (flow tip) sizes for smooth analysis and sorting of different cell types. The different nozzles should be with fixed alignment. · There should be provision of built-in or GMP compatible sterile sorting. · There should be provision of sort collection at low temperature. Thus instrument must have temperature control at both the ends (sample side/sort side) · During any sort interruption, there should be a provision for sample protection. · Should Zoom into close populations and give greater separation of each with ability to sort at that resolution. · The system should have the biohazard containment system. · Data management system: PC workstation with at least the following configuration - Pentium Duo Core @ 2.8 GHz, processor, Intel motherboard 1GB RAM, 160 GB Hard Drive, DVD/CD-ROM write/rewrite Combo Drive, Zip drive, Ethernet, two 19" LCD Color monitors, high quality colour laser printer. · The system should include software, which can be loaded onto a similar system (PC window based) for off line analysis. · All accessories such as a Starter Kit and UPS for full system must be provided. · Power: 220-240VAC 50 Hz · The company should have proven capability demonstrated in the past in after sales service and application support in Flow- cytometry instrumentation in India

· The company will provide free training of at least 2 personnel at manufacturing site in addition to on site training · The equipment must be provided with ergonomic Stand/Table for instrument and computers and a chair for the operator · Warranty of a minimum of 3 years has to be provided.

TENDER NO. PGI/MM/SGE/16/06-07

Recording System Specifications · Computer-A · Computer-B · Printer · Mobile Computer Table · Digital Camera and Tripod stand Specifications for Desktop Computer-A

· · · · · · · · Intel Core 2 Duo processor (2.66GHz, 2MB Cache) with Asus/Intel motherboard 2GB DDR2-RAM (2x1024) 320GB (160GB x 2) 7200 rpm SATA hard drive 18X DVD-R/RW drive AND Combo DVD-ROM/ CD-R/RW drive Premium CD/DVD Burner Software Multi-memory card reader 3 USB ports (all USB 2.0 compatible) on the front of the cabinet for easy access FireWire port (IEEE-1394) with FireWire cable (2 meters in length) and compatible software for downloading videos from commercial analog/digital camera/camcorders. LAN card (high performance) and 56K data/fax modem Graphics card ­Latest (at least 256MB memory) NVIDIA GeForce-8 series Sound card ­ Latest sound blaster card Logitech Multimedia Keyboard AND Logitech Optical Mouse Latest Pinnacle TV tuner/video capture card (preferably PCTV Pro PCI) 17-inch LCD Flat Panel Monitor Intex-2.1 Speakers, 2000watts PMPO Preferably including External Hard disk drive 100GB Windows VistaTM Home Premium Microsoft Office Home Edition 2007 Latest antivirus software (Kaspersky/McAfee) Premium high performance archival software for storing, editing, sorting still images and videos. Software should allow proper cataloging as well as easy retrieval of images/videos. Should be easy to work and customizable. All hardware, software and drivers should be Windows Vista compatible UPS with standby time of at least 30 minutes.

· · · · · · · · · · · · · ·

Specifications for Desktop Computer-B

· Intel Core 2 Duo processor (2.66GHz) with Asus/Intel motherboard

· · · · · · · · · · · · · · · · · ·

1GB DDR2-RAM 160GB 7200 rpm SATA hard drive 18X DVD-R/RW drive Premium CD/DVD Burner Software FireWire port (IEEE-1394) with FireWire cable (2 meters in length) and compatible software for downloading videos from commercial analog/digital camera/camcorders.. Multi-memory card reader Graphics and Sound card Pinnacle TV tuner/video capture card 3 USB ports (all USB 2.0 compatible) on the front of the cabinet for easy access Logitech Multimedia Keyboard AND Logitech Optical Mouse 17 inch LCD monitor Intex-2.1 Speakers, 2000watts PMPO Windows VistaTM Home Premium Microsoft Office Home Edition 2007 Latest antivirus software (Kaspersky/McAfee) Premium high performance archival software for storing, editing, sorting still images and videos. Software should allow proper cataloging as well as easy retrieval of images/videos. Should be easy to work and customizable. All hardware, software and drivers should be Windows Vista compatible UPS with standby time of at least 30 minutes.

Specifications for Printer

· · · · · Inkjet printer with separate back and color cartridges Should have functions of printing, copying and scanning Printing speed (maximum) should be 20 pages per minute (color) and 22 pages per minute (black), or more. Should have high quality and economy modes Scanning resolution (maximum) should be 1200x2400dpi (or more) optical resolution

Specifications for Mobile Computer Table for keeping the computer and printer

· · Side-to-side Width should be 60cm and Front-to-back Width should be 40cm Height should be designed to be functional as a computer table when sitting on a usual chair (with a sliding keyboard/mouse shelf) With good quality large wheels (with lock) for easy mobility by a single person Sturdy frame all around to protect the computer and monitor Space for Computer, Monitor, Printer, UPS, Keyboard and Mouse Space to carry all cables in the back. Should be designed for daily easy mobility and should be built of tough material to withstand daily shifting.

· · · · ·

Specifications of Digital Video Camera and Tripod stand

· · · · · · · · · · · · · Should have a recording media consisting of a hard disk drive of at least 40GB capacity Quality of recorded digital still and video should be at least 3 megapixel (gross) 16:9 aspect wide screen recording should be possible Zoom should be at least 10X optical and 20X digital zoom. Should have a built in microphone with audio recording in Dolby digital 5.1 ch surround sound quality. Should have at least 2.7 inch LCD screen. Should have a Memory stick DuoTM Media slot along with a compatible memory stick of at least 2 GB capacity. Should have an image stabilization system. USB 2.0 interface port with USB cable Should have a camera station cradle for battery charging and connecting to computer for image transfer and editing Should be compatible with stand alone DVD burners for DVD recording Compatible software for transfer of video and still images to computer and archiving Hard disk drive protection shock absorber to protect from physical damage during accidental drop or fall

·

Wireless remote with all important functions and battery included Extra-long life lithium battery (7.2V, 17.7 Wh/2460 mAh) with charger Spare lithium battery similar to above AC adapter Should include Audio-visual cables for transfer of files to computer Should include remote control tripod including pan handle and smooth function. It should expand to a maximum height of at least 55 inches. Should support a camera weight of at least 2 kg. Should include a semi-soft multipurpose camera case with compartments for extra-battery, chargers, cables and small accessories.

· ·

· · · ·

TENDER NO: PGI/MM/SGE/17/06-07

Surgical Instruments - Open Surgery Sl.No. Name of Instruments 1 BP Handle (No. 3) 2 BP Handle (No. 4) 3 BP Handle (No. 7) 4 Artery F. straight 160mm 5 Artery F. curved 160mm 6 Artery F. straignt Masq. 125 mm 7 Artery F. mosq.curved 125mm 8 Artery F.long 230mm 9 Artery roberts 230mm 10 Alice F. 150mm 11 Babcock 165 mm 12 Sponge holder (straight) 240 mm 13 Sponge holder (curved) 240 mm 14 Forceps diathermy long Cushing 250 mm 15 Forceps diathermy medium Cushing 200mm 16 Forceps packing long heavy standard 250mm 17 Forceps thumb medium heavy 200mm 18 Forceps tooth heavy 200mm 19 Lahey's heavy Desecting F. 230mm 20 Lahey's medium sweet 190mm 21 Lahey's fine Baby Adson 180mm 22 Czerny 23 Langanback 215mm (Kocher) 24 Morris ret. 240mm (50x55mm, 45x55mm) 25 Deaver ret. Large 300mm 26 Deaver ret, med. 215mm 27 Miculicz Abdo Ret 260mmx135mm 28 Miculicz Abdo. Ret 260mmx120mm 29 Self ret. (Balfoor) 260x200x90mm 30 Needle Holder long fine 230mm Quantity 1 1 1 6 6 6 6 4 2 4 4 2 2 2 2 2 2 2 2 2 2 2 2 2 1 1 1 1 1 2

31 Needle Holder medium fine 180mm 32 Needle Holder small fine 150mm 33 Needle Holder long heavy 200mm 34 Towel Cluips (Backhaus) 13cm 35 Sump suction 225mm 36 Fine tip suction 180mm 37 Little wood 38 Stone holding F. 220mm 39 Stone holding F. 200mm 40 Kidney tray small size 41 Bowl (all sizes) 42 Sargent Abd. Spatula 43 Scossprs (metzen) long 250mm 44 Scissors (metzen) medium 200mm 45 Scissors (metzen) small 180mm 46 Yaunkur suction tip 285mm 47 Table mounted self retaining abdominal Other Instruments 1 Protoscope 50x27mm 2 Rectal biopsy forceps 380mm 3 Hemo Clip applicator LT200 Surgical Instruments (Laparoscopic) 1 Veress Needle 120mm 2 Trocar Cannula 5.5mm Magnetic Sprial 3 Trocar Cannua 10mm Magnetic Plain 4 Washers 3.5mm Red 5 Washers 10mm Blue 6 Washers 5.5mm Orange 7 Reducer Metal 10 to 5 mm 8 Reducer Metal 5/3.5mm 9 Dissector Maryland short (5mm) 10 Dissector Maryland Long (5mm) 11 Atraumatic Graspers Plain 5mm 12 Atraumatic Graspers Endoclinch 5mm 13 Atraumatic Graspers Fenestrated small 5mm 14 Atraumatic Graspers Fenestrated Medium 5mm 15 Atraumatic Graspers Fenestrated long 5mm 16 Atraumatic Graspers Babcock Plate Type 10mm 17 Traumatic Graspers 2/3 Tooth 5mm 18 Traumatic Graspers 5mm Tenaculum 19 Scissors Curved Maryland 5mm 20 Clip Applicators 3 in 1(10mm), LT200/300/400 21 Needle Holder Strainght Jaws Tc ethicon type 5mm

2 2 2 6 2 1 1 1 1 2 5 1 1 2 2 1 2 2 2 2

1 3 3 3 5 5 2 2 2 1 1 1 1 1 1 2 1 1 2 1 1

22 Nedle Holder self Allgning 5mm ethicon type 23 Handles Fibre Handle Plain 24 Handles Metal Handle bottom Ratchet 25 Suction Irrigation 2-Way sliding Valve 5mm 26 Retractors Fan Three Prong 5mm 27 Retractors Fan Three Prong 10mm 28 Monopolar electrodes 'L' Hook 29 Port Closure Storz Type 30 Palanivelu Hydatid Cannula

1 2 2 1 1 1 2 1 2

TENDER NO. PGI/MM/SGE/18/06-07

Technical specification for Diathermy

Facility Monopolar with a. cut b. soft coagulation c. forced coagulation d. spray coagulation Biopolar with a. cut b. coagulation

Monopolar cut · wave form of voltage : sine wave · normal frequency : 330 kHz · blend setting : adjustable in > 8 settings · max. output : 300 W in increment of 01 watt · activation : by finger/foot switch · acoustic tone : continuous · output power indication :+ · with spark regulation & permanent initial cut support. Monopolar soft coagulation · Regulated coagulation without sparking. · Wave form of voltage : sine wave · Nominal frequency : 330 kHz · Max. output power : 120 W at 100 to 200 Ohms · Activation : finger & foot switch · Acoustic Signal :+ · Stopping power delivery : auto and manual Monopolar forced coagulation · with spark control & spark regulation · wave form of voltage : modulated pulse · nominal frequency : 1 MHz

· · · · ·

max. output power max. peak voltage power increment Activation Acoustic tone

: 120W at 300 ohms : 3200 Vp : 1 watt : finger & foot switch :+

Monopolar spray coagulation · with spark control & regulation · wave form of voltage : modulated pulse · nominal frequency : 1 Mhz · max. output power : 120 W with increments of 1 watt · max. peak voltage : 5700 Vp · Activation : by finger & foot switch Bipolar cutting · with permanent initial cut support · with spark coagulation · wave form of voltage : sine · nominal frequency : 330 kHz · Blend setting : in >8 setting · Max. power output : 150 W with increments of 1 watt · Autostart delay & manual setting : + · Activation : foot switch · Acoustic signal :+ Bipolar coagulation · regulated coagulation without sparking · wave form of voltage : sine wave · nominal frequency : 330 kHz · max. power output : 120 W with increment of 1 watt · max. peak current : 2A at 75 Ohms · Activation : autostart with delay /foot switch · Acoustic signal :+ Bipolar high current coagulation "Ligation" · without sparking · wave form of voltage : sine wave · nominal frequency : 330 kHz · max. output power : 200 W at increment of 1 watt · max. peak current : 4A at 10 Ohms · Activation : by foot/finger switch Programes 70 to 100 memory with min 20 preset for laparoscopic surgery, TUR, heart surgery, microsurgery, ligation and general surgery. Safety · Class I · Type CF according to IEC601-1/2 · Permanent leakage current control · Integrated safety system check · Dosage rate control · Permanent self check Documentation · detection and storage of error messages · operating mistakes · current modes and current intervals

· external display with PC adaptation Input characteristics · Voltage : 230 Vac + 15%-20%, 50Hz · Line fuse : 5 A slow-blow · Idle power consumption : 40 VA maximum · Maximum power consumption : 800 VA maximum - Portable - Accessories

TENDER NO. PGI/MM/SGE/19/06-07

Technical specifications for Pneumatic compression device

· · · · · · Compression type: Sequential, gradient and circumferential pressure. Compression cycle: Having individualised vascular refill detection Alarms: Audiovisual for low or high pressure and for any malfunction Have sleeve cooling function and a cycle monitor. Attachable to the bed with adjustable height and width Accessories including knee length, small thigh length, medium thigh length and large thigh length sleeves and knee length and thigh length stockings

TENDER NO. PGI/MM/SGE/20/06-07

Specifications of scanner (X ray and Documents)

· · · · · · · · · Should be a flat bed scanner Should be able to capture X-ray image resolution of medical diagnostic grade. The best optical image resolution should be at least 1600x3200dpi (Preferably 3200x6400 dpi). Should be able to scan large X-ray film size of upto 12x16 inches in one go. Scanning speed should be 60 seconds (or less) for a large film of 12x16 inches size. Should be able to capture and store images files in all conventional formats such as JPEG, GIF formats. Should also provide a 12 bit grayscale image and should be convertible to color scanning modes if required. Should have the requisite software for image recording, archiving, editing and retrieval. The required power supply should be compatible with the standard Indian electrical connections or else the required adapters should be included.