PHILIPS/HP® SERIES 50A ANTEPARTUM FETAL MONITOR

· · · Advanced fetal monitoring features Accurate and continuous ultrasound FHR traces Ideal for clinics and private doctors office

Typical Manufacturer's Picture

Specifications

Monitor Power Requirements Operating Voltage: 100V - 120Vor 220V - 240V (±10%). Line Frequency: 50 to 60Hz ±5%. Power Consumption: 25VA max. Battery Type: 2 x 1.5V (AA size). Lifetime > 1 year. Weight and Dimensions Height: 115mm (4.5in). Width: 340mm (13.4in). Depth: 308mm (12.1in). Weight: 5.74kg (12.6lb) (without transducers). Displays Numerical Display M1351A Single Ultrasound Model: One heart rate display (orange) and one uterine activity display (green). M1351A Dual Ultrasound Twins Model:Two heart rate displays and one uterine activity display. M1353A Model: Two heart rate displays and one uterine activity display. Type: 7-segment LEDs (10mm). FHR Range: 50 to 240 bpm. Uterine Activity Range: -99 to +127 relative units. Instrument Display Telemetry Mode is displayed if Option J10 is fitted and an or M1310A or 80240A Fetal Telemetry System is connected and switched on. M1351A Single Ultrasound Model: One signal quality indicator. M1351A Dual Ultrasound Twins Model:Two signal quality indicators. M1353A Model: Two signal quality indicators. Inputs M1351A Single Ultrasound Model US socket accepts the M1356A ultrasound transducer. Toco socket accepts the M1355A Toco transducer. Socket for the Remote Event Marker (15249A), and another for servicing. The monitor automatically selects the correct operating mode. Ultrasound Mode System: Pulsed Doppler oscillation. Frequency: 998.4 kHz. Repetition Rate: 3.2 kHz. Ultrasound Intensity: 1.5mW/cm² average for each of the seven active surfaces. DECG and MECG

166 Highland Park Dr. Bloomfield, CT 06002 U.S.A. Tel: (860) 218-2575 Fax: (860) 218-2565 Email: [email protected] Rev 1 1/05

External Labor Signal Range: 0 to 127 units. Offset Compensation: ±200 units. Internal Labor Signal Range: -99 to +127 mmHg. Patient Leakage Current: 10 µArms. Sensitivity: 40 µV/V/mmHg (M1348A). 5 µV/V/mmHg (M1334A and CPJ840J5). Recorder Mechanism: 3-channel, high-resolution (8 dots/mm) thermal array recorder with paper-end detection. Paper Speeds: 1, 2 or 3 cm/min. Recording Time Per Packof Paper: 1 cm/min (25 h).2 cm/min (12 h 30 min). 3 cm/min (8 h 20 min). Paper Advance Speed: 24 cm/min (with automatic stop at the paper-end mark). Annotation: Time of day, date, and paper speed are printed automatically every ten minutes. Monitoring mode is printed with every alteration of parameter. Paper: Fanfold paper with numbered pages. FHR Scale: USA: 30 to 240 bpm @ 30 bpm/cm. Other countries: 50 to 210 bpm @ 20 bpm/cm. Labor Scale: 0 to 100 units @ 25 units/cm. Self-Test Facilities Self-test facilities include: System test: With no transducers connected (includes a display and recorder test). Parameter test: With the appropriate transducer connected, the monitoring mode (ultrasound or uterine activity) is tested. Combined Interface Module Telemetry: M1310A Fetal Telemetry System. System: M1383A/B/C OB TraceVue. Either Barcode Reader: SmartWand. or Maternal NIBP Monitor: Dinamap 1846/8100. COLIN Press-Mate /Nippon Colin Listmini Model BP-8800. or Nellcor OxiFirst Fetal Oxygen Saturation Monitor (N-400) or compatible. Modem Interface Module Modem: Interface socket for an Philips-approved PCMCIA card modem. Fetal Trace Memory: Local fetal trace storage. Barcode Reader: Smart Wand. RS232 Serial Interface: For internal use only. Remote Event Marker (15249A) Length: 2.8m/9ft 2in. Weight: 75g/2.65oz.

Soma Technology, Inc. acknowledges all registered trademarks of manufacturers' listed. The technical data given in this publication are for general information and are subject to change without notice.

166 Highland Park Dr. Bloomfield, CT 06002 U.S.A. Tel: (860) 218-2575 Fax: (860) 218-2565 Email: [email protected] Rev 1 1/05

166 Highland Park Dr. Bloomfield, CT 06002 U.S.A. Tel: (860) 218-2575 Fax: (860) 218-2565 Email: [email protected] Rev 1 1/05

Information

Blank Format

3 pages

Find more like this

Report File (DMCA)

Our content is added by our users. We aim to remove reported files within 1 working day. Please use this link to notify us:

Report this file as copyright or inappropriate

607386


You might also be interested in

BETA
Layout 1
Blank Format
2000947-004 C.book