Technical Description and Operating Instructions

Röntgen-GammaDosimeter 27091

September 02, 2008

Contents

1 2 3 3.1 3.2 4 4.1 4.2 4.3 4.4 4.5 4.6 4.7 5 6 6.1 6.2 6.3 7 8 9 10 Applications Safety Precautions System Components Scope of Delivery Accessories Using Guide Preparing and Battery Changing Outlook and Functions of operational Controls, Display and Connectors Check of Display, Battery Voltage and electrical Zero Adjustment Correction of Measurements Adjustment of Alert Levels Measuring Range Selection Usage of the Wall Reinforcement Cap Usage of Auxiliary Devices Service and Maintenance General Precautions Decontamination Drying of the Measuring Probe Storage Conditions Transport Conditions Technical specification Internal Structure 3 4 5 5 5 6 6 8 11 14 14 16 19 20 21 21 21 22 22 23 23 26 27

Appendix

2

1

Applications

The X-Ray-Gamma-Dosimeter 27091 is the new designed successor of the well-known dosimeter VA-J-15, 27040 and 27060. It is a battery-operated dosimeter and dose rate meter provided with an ionization chamber. It will be used in nuclear engineering laboratories, nuclear medicine clinics, radiation facilities and reactor units for measuring continuous roentgen and gamma rays. This device is applicable for measuring of pulsed roentgen rays in the dose rate mode also. Its high measuring sensitivity and the low dependence on energy and direction of the probe allow for reliable measurements. Besides its main application as a precision radiation protection dosimeter, the device can also be used as a dosimeter for high dose rates thanks to its wide measuring range. For such applications, the measuring probe can be located up to 100 m from the display unit. Beta radiation from energies of E 160 keV can be qualitatively detected. It is possible to use the dosimeter in a wide field of applications. You can take it in different kinds of applications, so as a mobile or as a stationary device, if it required. Mobile applications

If it is necessary to measure the doses or dose rates at many points in your fields, a carrier bag is right to do it easily. The bag is one part of the delivery. Stationary applications Remove th handle and place the device on stable and level base!

3

Remove the Measuring Probe and the Handle and insert both Blinds after connecting the Display Unit with the probe by an extension cord. The maximum cable length can range up to 100 m.

2

-

Safety Precautions

The Ion Chamber is the "Heart" of your dosimeter 27091. Handle with care ! The low areal density forbid any mechanical stress ! Don't transport the Measuring Probe or the complete dosimeter without placed Wall Reinforcement Cap ! If you must transport the device, switch the dosimeter out and put the Basic Range Control in the Zero Point Adjustment position ! Never try to repair the dosimeter by yourself ! There are no user-serviceable parts inside !

-

Attention! Don't touch the Measuring Probe, if the Ion-Chamber is destroyed or removed and the dosimeter is switched on ! It is possible to contact high voltage up to 500 V !

4

3

System Components

3.1 Scope of Delivery After unpacking the shipping carton, check if you received the following items: 1) Measuring Probe with Wall Reinforcement Cap 2) Display Unit 3) Mountable Handle 4) Set of Operating Instructions 5) Carrier Bag with shoulder strap 6) 1.5 m Connecting Cable 7) Case for Accessories 8) 2 Blinds

3.2 Accessories If you need a long distance between your points and the Display Unit, you can use an extension cord. Following length are possible. Please ask your dealer. 15 m Extension Cord 50 m Extension Cord It's possible to deliver different cable length, if you wish it.

5

4

Using Guide

4.1 Preparing and Battery Changing The dosimeter is delivered without batteries. You must insert 4 round cells into the device before starting the first measurement. Loading the batteries 1. Unpack the device out of the transport case

2. Put the dosimeter with upper side Down on a stable and level base.

3. Push the bottoms at right and left side into the case and pull-up this part.

4. Place the parts after another.

5. Unscrew the Measuring Probe.

6. Insert the batteries, aligning there Polarities as indicated inside at the Nickel coated clamps.

6

7. Slide the first cell to the front side and push the next into the clamp.

8. Repeat this at the next side.

9. Place the mechanical connection at the Basic Range Control of the probe correctly. Screw the Measuring Probe at the connector.

10. Insert the part into the part beginning with the backside snapper.

11. Push both parts together careful.

7

Battery Changing Do the same like the first preparing, but remove the cells by doing the step 6 in the reverse order. Attention! Load new batteries with there aligned polarities correctly ! When the dosimeter is not in use for long time, remove the batteries and store them in dry and cool room ! Remove spent batteries immediately ! Protect the Enviroment !

Do not discard used batteries with your household rubbish !

4.2 Outlook and Functions of operational Controls, Display and Connectors 1. Measuring Probe with Wall Reinforcement Capp Remove the cap: - Turn to right - Pull to the front side

2. Display Unit Part 1 (Upper part) 3. Display Unit Part 2 (Under part) 4. Handle with Toogle Fastening Remove the Handle: - Turn the toogle to right - Pull up the toogle 5. BNC-Jack for recorder connection

8

6. LCD Screen

Selected Alert Level Digital Read Out Selected Basic Range Selected Final Value Horn Symbol (Blinking, if the measured level is higher as the selected threshold) Quasi Analogue Bar Chart Battery Symbol indicated Undervoltage 7. Front Panel of the Display Unit Part 1 (Upper part) Basic Range Control

Piezo Buzzer

9

8. Front Panel of the Display Unit Part 2 (Under part) Light

Control Zero Point Adjustment Control Final Value Select

10

10. Measuring Probe

Round Connector Basic Range Control

4.3 Check of Display, Battery Voltage and electrical Zero Adjustment Display Check: 1. Switch the device ON by the control: Final Value Select Position : 2000

2. All Segments of the LCD will be active for few seconds.

3. Compare the read out with the picture Above.

11

4. Please consult your dealer or the manufacture, if any error detected. Check of the Service Voltage: The batteries are been always watched by the device inside. In the case of undervoltage a blinking battery symbol indicates these conditions.

Change the batteries immediately !

Zero Point Adjustment: Attention! Do this: Check and adjust the electrical Zero Point of the input stage before every measurement Do this also, if environmental temperature is changed !

1. Switch Power On by the Final Range Control. Position: 2000

2. Wait 2 minutes !

Power On Delay Time !

12

3. Select the Basic Range Position:

0

4. Select the Final Value Position : 20

5. Use the Control Zero Point Adjustment to adjust the read out at a value of 0.05 !

6. The first segment of the Bar chart could be visible !

13

4.4 Correction of Measurements The dosimeter is calibrated at Refernece Conditions (see Technical Data) by the manufacture. Perhaps, the physical data of the ion chamber dependence on air pressure and temperature. Correct the dependence on Air Pressure p and Temperature use therefore the following equation or the diagram in the appendix. K0 = K f

f =

(273 + [°C ]) 101,33 293 p[kPa]

If the Photon Energy of radiation is known approximately, it is possible to reduce the error of energy dependence by calculation. Use therefore the diagram in the appendix to get the Sensitive on Energy q/q0 of the dosimeter and calculate:

Venergy = V *

These methods reduced the Additional Error. Energy Range 6 keV ­ 15 keV 15 keV ­ 2 MeV 4.5 Adjustment of Alert Levels 1. Unpack the device out of the Transport Case. 2. Put the dosimeter with upper Side down on a stable and level base.

q0 q

V: Read out without correction

Reduced Additional Error ± 10 % ±5 %

14

3. Push the bottoms at right and left side into the case and pull-up this part.

4. Place the parts after another.

5. Insert the part 2 into the part 1 beginning with the backside snapper.

15

6. Push both parts together careful.

4.6 Measuring Range Selection The wide measuring range of the dosimeter is split in 3 Basic Ranges and each of them is also split in 3 linear fields.

Attention! Always atart with the most insensitivity Final Value (2000) ! If you change the Basic Range, select the Final Value 2000 also before !

16

Dose rate measuring: If it not possible to estimate the heigh of dose rate, start in the most insensitive range. If the read out is lower Than 10 % of the Final Value, change the final range. If the read out in the lowest final range is also lower than 10 % of the Final Value, do this:

1. Select the Final Value 2000

2. Select the Basic Range Position:

0

3. Wait ca. 120 seconds.

Time is required for discharging of the Ion Chamber and the input stage.

4. Change the Basic Range.

If the read out is lower than 10 % of the Final Value, change the final range.

17

Dose measuring:

1. Select the Basic Range Position:

0

2. Select the Final Value 2000

3. Change Basic Range for a short time into Position: mSv/h

Time is required for discharging of the Integration capacitor of the Input stage.

4. Change the Basic Range Into position: µSv The measuring period starts In this moment.

If the read out is lower than 10 % of the Final Value at the end of your measuring period, change the final range.

The new measuring period starts after repeating the change of the Basic Range into Position µSv.

18

Note: For high exactness watch the read out in a period of 1 minute and calculate the average:

K=

-

(Vmin + Vmax ) 2

In the most sensitivity range of dose and dose rate it is possible to correct the influence of leakage currents of the probe by following calculation: Dose: Read Out ­ 0.01 µSv / min Dose Rate: Read Out ­ 0.6 µSv / h

-

If the environmental conditions had been changed by transport, wait about 2 hours. In this time the temperature inside of the probe will adapt to the environmental temperature.

4.7 Usage of the Wall Reinforcement Cap The large Energy Range of the Dosimeter required the placement or the displacement of the Wall Reinforcement Cap in dependence on the kind of radiation; X-Rays, gamma radiation or Bremsstrahlung. Radiation x-rays, x-rays, x-rays, Energy 6 ­ 100 keV 20 keV ­ 3 MeV > 3 MeV Wall Reinforcement Cap NO Yes Yes + 2 cm thick PMMA plate NO Direction of radiation must be vertical to surface of the Measuring Probe Qualitatively only Note

Do this:

160 keV

19

1. Switch the Power On by the Final Range Control Position: 2000

2. Select the Basic Range Position:

0

3. Remove the Wall Reinforcement Cap Turn to right Pull up the cap carfully

Do this in reverse order for placement of the cap. Attention! Always select the Basic Range 0 Position!

5

Usage of Auxiliary Devices

Extension Cords: The Measuring Probe can be located at a distance of up to 100 m from the Display Unit. Use an Extension Cord for this application. In addition to the standard connection cable, you can order Extension Cords with 15 m, 25 m, 50 m or different length.

20

Attention! Do always: Switch the Power out, if you start to remove the Measuring Probe! Adjust the electrical Zero Point again, if you had connected the cable between Display Unit and the Probe. Correct the dependence on temperature and air pressure by calculation. Note: Only the standard connecting cable with a length of 1.5 m is a part of the delivery. Commercial Tripods The Measuring Probe is provided with a 3/8" thread. Thus can be used for fastern customary by trade tripod. Use a connecting piece fore this operation perhaps.

6

Service and Maintenance

6.1 General Precautions The Device RGD 27091 and 27091-U is a precision radiation protection dosimeter. For this application it is liable for gauging. There is prohibited to open the device. The Dosimeter is sealed. The user had only to do: Battery Changing Cleaning Function Check every 6 month

6.2 Decontamination It is necessary to clean the Display Unit, the Measuring Probe and the removed Wall Reinforcement Cap, if the device radioactive contaminated. You should use a cloth, wetted by soap suds for this. The cleaning of the Ion Chamber could not be successfully. The low areal density and the kind of surface is not enough resist to remove the contamination completely. Foe expecting contamination, use an effective protection for the Ion chamber, e.g. polyethylene bag or foil.

21

Attention! Don't use solvents, e.g. petrol, benzene or acetone for cleaning! In this case, you will damage the dosimeter!

6.3 Drying of the Measuring Probe If you transport the dosimeter from a location with temperature at the lowest limit to location with higher temperature and high air humidity, water will condensed at the device. This is a possible reason for increasing of leakage currents. The error in this case can reach up to > 1.0 µSv/h. Wait until the device is dry before starting any measurements. Use a heating compartment or an air blower to speed up the process, but the temperature must be lower then 50 °C! Remove the Wall Reinforcement Cap before. Attention! Don't touch the Measuring Probe, if the Ion Chamber is destroyed or removed and the dosimeter is switched ON! It is possible to contact high voltage up to 500 V!

Never try to repair the dosimeter by yourself! There are no user-serviceable parts inside!

7

Storage Conditions

-25 °C to +55 °C 800 hPa to 1100 hPa 80 %

- Storage temperature range - Outside air pressure - Relative air humidity

- when the dosimeter is not in use for long time, remove the batteries and store them in dry and cool room! - Pack the dosimeter into the Transport Case before storage them! - The storage room must be cool and dry, free of aggressive chemical or solvent vapors!

22

8

-

Transport Conditions

The Ion Chamber is the "Heart" of your dosimeter 27091. Handle with care ! The low areal density forbid any mechanical stress ! Don't transport the Measuring Probe or the complete dosimeter without placed Wall Reinforcement Cap ! If you must transport the device, switch the dosimeter out and put the Basic Range Control in the Zero Point Adjustment position ! Temperature range Outside air pressure Relative air humidity -25 °C to +55 °C 800 hPa to 1100 hPa 80 %

-

9

Units

Technical specification

Photon equivalent dose Photon equivalent dose rate

Read Out - Dose in 3 linear fields - Dose Rate basic range in 3 linear fields basic range in 3 linear fields Measuring Ranges In the final range 20 µSv/h, the measuring range starts at 3 µSv/h In all other indicating ranges, the measuring range starts at 10 % of the final value disabled or adjustable at 30 %, 60 % or 90 % of the final value Piezzo-buzzer Horn-symbol on LCD-Screen Both are independent ON/OFF switchable µSv final values: 20, 200 and 2000 µSv/h final values: 20, 200 and 2000 mSv/h final values: 20, 200 and 2000

Thresholds values/ Alert levels - acoustic Signal - optical Signal Energy ranges - without wall reinforcement cap - with wall reinforcement cap

6 keV to 100 keV 20 keV to 3 Mev

23

(7.5 MeV with additional PMMA shielding) Radiation Detector - volume - areal density Wall reinforcement cap - areal density Refernce point Overload resistance Reference Conditions - temperature - air pressure - photon energy - radiation direction - distance between the radiation source and the refernce point of ion chamber - correction of air pressure Gauging error in all rages Additional error - in depence on energy 6 keV to 15 keV 15 keV to 2 MeV 2 MeV to 7.5 MeV ±45° ±5 % ±90° ±10 % ±3 % / 10 K -5 % +1 µSv/h cancelled -40 % ±10 % ±30 % 20 °C 101,3 kPa 662 keV (Cs-137) radial 0,8 m done max. ± 15 % under reference condition air equivalent ionization chamber 600 cm³ 35 mg/cm² removable 500 mg/cm² Centre of the ion-chamber at a distance of 6 cm from the front side 100 Sv/h in all ranges

- directional dependence of relative measuring sensitivity - dependence on temperature (constant air pressure) - saturation deficit at 2000 mSv/h - error by leakage current - zero point error after zero point adjustment

24

Time characteristics - delay after power-on or basic range changing - set-up time (from start to 95 % of the final value in the dose rate mode) range 20 µSv/h all other ranges Check instruments - performance check and calibration Recorder connection by means of BNC jack Power supply Continuos operation without changing of batteries Working temperature range Storage temperature range Outside air pressure Relative air humidity Degree of protection - Display Unit - Measuring Probe with wall reinforcement cap Dimensions - Display unit (length x widths x heights) - Measuring Probe (diameter x heights) Weight Display unit with measuring probe 2.5 kg 210 x 120 x 135 mm³ 112 mm x 260 mm IP 40 IP 40 blinking battery symbol in case of undervoltage 10 mV for 100 Ohms if measuring range final value is indicated 4 round batteries LR 14 (DIN IEC 86 T1) or 4 NiCd round cells R 14 (DIN 40729) 150 h -10 °C to +45 °C -25 °C to +55 °C 800 hPa to 1100 hPa 80 % 120 s

20 s 8s

25

Mechanical stability - test conditions DIN IEC 68 part 2-6 DIN IEC 68 part 2-29 (10-500) Hz, 0.15 mm, 2 g peak acceleration: 150 m/s² shock period: 6 ms shock counts: 500

10

Internal Structure

A given radiation field, with a photon equivalent dose rate HX generate a current IC in the ion chamber. The volume of the ion-chamber is V. The factor represents the actually air density in dependence on air pressure and temperature. The value of IK is the result of the following equation in the case of saturation:

I C = K1 * * V * H X

K1 ... proportional factor The voltage at the chamber UC is high enough to minimize the recombination of charges in the volume of the chamber.

The current IC flow into the inverting input of the amplifier A. All basic ranges are controlled by the switch S1. The final value is selected by R2. The output voltage of the amplifier is 1 * R1 * I C A in the dose rate mode and

UA = -

UA = -

1 * I C * dt A*C

in the dose mode.

26

The variable A is a result of a factor K2 multiplied with the final value of the selected range.

A = K2 * F

where: F = 2000, 200 or 20 (final value) The analogue to digital converter ADC converts the voltage UADC to proportional digital value. This value is shown as a digital value and as a quasi analogue bar chart at the LCD. The potentiometer R4 "Correction" will be used to compensate influences of air pressure and temperature. The offset-voltage of the input stage of the amplifier A are adjusted by the control "zero point adjustment". The switches S3 and S4 allow to set or reset the optical and the acoustic alert levels. All supply voltages are generated by a transverter.

Appendix

Diagramm 1: "Correction Factor in Dependence on Air Pressure and Temperature" Diagramm 2: "Dependence on Energy of RGD 27091"

27

28

29

Information

Microsoft Word - RGD 27091 - Gebrauchsanweisung .doc

30 pages

Report File (DMCA)

Our content is added by our users. We aim to remove reported files within 1 working day. Please use this link to notify us:

Report this file as copyright or inappropriate

1154269


You might also be interested in

BETA
Microsoft Word - RGD 27091 - Gebrauchsanweisung .doc
VD DAP Kermax plus TinO QAe 002