Our Thermo Scientific Forma Class II, Type B2 cabinet provides a clean air environment with maximum product, personnel, and environment protection from agents assigned to biosafety levels 1, 2, or 3. Total exhaust makes the B2 ideal for work when volatile toxic chemicals and radionuclides are required as an adjunct to microbiological studies.

Product Specifications

Thermo Scientific Forma Class II, Type B2 Biological Safety Cabinet

Maximum Protection, Customer Conveniences

· Cabinet's 12.0" (305mm) supply air intake and exhaust ducts include adjustable, zero leak damper for convenient connection to your in-house systems. · Service valves can be plumbed from the side or rear of the cabinet for convenience, and all maintenance procedures can be performed from the front of the unit. · Pressure differential switch continuously monitors the exhaust pressure for safety.

In the event of inadequate exhaust, the pressure switch interrupts power to the supply blower/motor, preventing contaminated air from blowing onto you or into the lab. · Exhaust failure alarm includes normally open, normally closed, and common remote alarm contacts to provide an output signal for remote monitoring capabilities via your in-house central alarm systems.

Forma Class II, Type B2 Biological Safety Cabinets ­ the ulitmate safeguard

Part of Thermo Fisher Scientific

Specifications and Ordering Information

Total Exhaust, Total Safety Room air is pulled into the cabinet by the blower and pushed through the supply HEPA filter, providing clean unidirectional air onto the work surface. Approximately 50% of this work area air exits through the rear grille. The remaining 50% meets room air and is pulled into the front intake grille, providing a protective barrier. The divided downflow and inflow air then combines and is pulled up the rear plenum through the exhaust HEPA filter by the facility exhaust system.

Optional Adjustable Modular Stand Stand is designed in accordance with American Disabilities Act (ADA) guidelines. Work surface height is adjustable in 1.0" (25mm) increments. Stock No. 191580 191581 Description For 4 Ft. B2 Units For 6 Ft. B2 Units

Ambient Air HEPA Filter Contaminated Air (negative pressure) HEPA Filtered Air

Thermo Scientific Forma Class II, Type B2 Biological Safety Cabinet

Specifications Models 4 Ft. Benchtop 6 Ft. Benchtop Work Area Width ­ 1149 Width ­ 1169 Height Depth Sash Opening Exterior Dimensions Width ­ 1149 Width ­ 1169 Height Depth Electrical, Plug, Breaker & Motor 1149 1169 Plug Breaker Requirement Motor Exhaust Requirements1 1149 1169 1149 1169 46.5" (1181mm) 70.5" (1791mm) 25.3" (643mm) 25.0" (635mm) 8.0" (203mm) 50.5" (1283mm) 74.5" (1892mm) 69.8" (1773mm) 35.6" (904mm), deduct 3.1" (94mm) for light canopy 220V, 50Hz 220V, 50Hz CEE 7/7 15 Amp 1149: 1/3 HP 1169: 1/2 HP 836-945 CFM2 1165-1332 CFM2 Exhaust Requirements (continued) Connection Service Valves 1149 1169 Additional Plugged Penetrations 1149 1169 Outlets 230 Unit Heat Load 1149 1169 Weight Net 1149 1169 Shipping (Motor) 1149 1169

1

12.0" (305mm) diameter duct connection, 1.5" to 2.0" w.g. psi drop at the connection 1, right sidewall 2, 1 on each sidewall 3, 1 on right and 2 on left sidewall 2, 1 on each sidewall CEE 7, 1 on each side wall (12 Amps total) 370 BTUH (110 Watts) 590 BTUH (170 Watts)

604 lbs. (274.0 kg) 904 lbs. (410.1 kg) 750 lbs. (340.2 kg) 1,100 lbs. (499.0 kg)

Units require large volumes of supply and exhaust air, critical to the proper operation of the Class II, B2 cabinets. It is the user's responsibility to supply the correct volumes of make-up air. Cubic Feet per Minute (CFM) is based on a Pitot tube in the exhaust duct, under clean conditions. 2 Cubic feet per minute (CFM) is based on the direct airflow method.

Height Stand Height Work Surface Height Total Height of Cabinet with Stand1

1

Minimum Height (Leveler In) 23.4" (594mm) 26.5" (673mm) 93.2" (2367mm)

Maximum Height (Leveler Out) 32.7" (831mm) 35.8" (909mm) 102.5" (2604mm)

Includes the height of the exhaust duct

North America: USA/Canada +1 866 984 3766 Europe: Austria +43 1 801 40 0, Belgium +32 2 482 30 30, Finland +358 9 329 100, France +33 2 2803 2000, Germany national toll free 08001-536 376, Germany international +49 6184 90 6940, Italy +39 02 95059 1, Netherlands +31 76 571 4440, Russia/CIS +7 (812) 703 42 15, Spain/Portugal +34 93 223 09 18, Switzerland +41 44 454 12 12, UK/Ireland +44 870 609 9203 Asia: China +86 21 6865 4588 or +86 10 5850 3588, India +91 22 5542 9494, Japan +81 45 453 9220, Other Asian countries +852 2885 4613 Countries not listed: +49 6184 90 6940 or +33 2 2803 2000

© 2007 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

www.thermo.com/bsc

BRO-LECAFBSCB2-0807

Information

BRO-LECAFBSCB2-0807.qxd

2 pages

Report File (DMCA)

Our content is added by our users. We aim to remove reported files within 1 working day. Please use this link to notify us:

Report this file as copyright or inappropriate

541767


You might also be interested in

BETA
DE
84077 Franklin 3:25.5 x 19
BRO-LECAFBSCB2-0807.qxd